191
A Technical Data
Lamb Wave Transducers (for ATEX Zone 1)
technical type
GRH1N33
GRK1N33
GRG1N33
order code
GLH-NA1TS
GLK-NA1TS
GLG-NA1TS
transducer frequency MHz 0.3
0.5
0.2
medium pressure
1
min. extended
bar
metal pipe: 10
metal pipe:
10 (> DN 120)
5 (< DN 120)
metal pipe: 10
min.
bar
metal pipe: 15
plastic pipe: 1
metal pipe:
15 (> DN 120)
10 (< DN 120)
plastic pipe: 1
metal pipe: 15
plastic pipe: 1
outer pipe diameter
2
min. extended
mm
120
60
190
min. recommended
mm
140
80
220
max. recommended
mm
600
300
900
max. extended
mm
1000
500
1600
pipe wall thickness
min.
mm
7
4
11
max.
mm
15
9
23
material
housing PPSU
withwith
stainless steel
cap304
(1.4301)
PPSU with
stainless steel
cap304
(1.4301)
PPSU with
stainless steel
cap304
(1.4301)
contact surface
PPSU
PPSU
PPSU
degree of protection
according to EN 60529
IP 65
IP 65
IP 65
transducer cable
type 2549
2549
2549
length m
5
5
5
dimensions
length l
mm
128.5
128.5
128.5
width b
mm
50
50
50
height h
mm
67.5
67.5
67.5
dimensional drawing
operating temperature
min.
°C
-40
-40
-40
max.
°C
+170
+170
+170
continued on next page
h
b
l
h
b
l
h
b
l
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......