19 Troubleshooting
19
Troubleshooting
If any problem appears which can not be solved with the help of this manual, contact our
sales office giving a precise description of the problem. Specify the type, serial number
and firmware version of the flowmeter.
Calibration
FLUXUS is a very reliable instrument. It is manufactured under strict quality control, us-
ing modern production techniques. If installed as recommended in an appropriate loca-
tion, used cautiously and taken care of conscientiously, no troubles should appear. The
flowmeter has been calibrated at the factory and usually, a re-calibration of the flowmeter
will not be necessary. A re-calibration is recommended if
• the contact surface of the transducers show visible wear or
• the transducers were used for a prolonged period at a high temperature (several
months >130 °C for normal transducers or > 200 °C for high temperature transducers).
The flowmeter has to be sent to FLEXIM for recalibration under reference conditions.
The display does not work at all or always fails
Make sure that the correct voltage is available at the terminals. The voltage is indicated
on the metal plate below the outer right terminal. If the power supply is ok, the transduc-
ers or an internal component of the flowmeter are defective. Transducers and flowmeter
have to be sent for repair to FLEXIM.
The message SYSTEM ERROR is displayed
Press BRK to return to the main menu.
If the message is displayed repeatedly, note the number in the lower line. Track down the
situations when the error is displayed. Contact FLEXIM.
The flowmeter does not react when key BRK is pressed during measurement
A program code has been defined. Press key C and enter the program code.
The backlight of the display does not light, but all other functions are available.
The backlight is defective. This problem has no influence on the other functions of the
display. Send the flowmeter to FLEXIM for repair.
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......