8 Installation of the Temperature Probe (Option)
8
Installation of the Temperature Probe (Option)
8.1
Fixing of the Temperature Probe
• Remove rust, insulation material and loose paint to guarantee a good thermal contact.
• Clean the pipe.
• Mount the protection plate and the isolation foam to the temperature probe (see Fig.
Fig. 8.1: Temperature probe:
a) with protection plate and isolation foam
b) without protection plate and isolation foam
- Apply a film of thermal conductivity paste (no FLEXIM supply) onto the contact sur-
face of the temperature probe.
- Take the spring end of the ball chain and insert the last ball in one of the slots on the
top of the temperature probe.
- Place the chain around the pipe. Pull the chain firmly and insert it in the second slot
of the temperature probe (see Fig. 8.2).
Fig. 8.2: Installed temperature probe
a)
b)
slot at the upper
side of the
temperature probe