18 Outputs
If the switching condition
QUANTITY
is selected in the scroll list
FUNC
, the limit of the out-
put has to be defined:
A positive limit will be compared to the totalizer value for the positive flow direction.
A negative limit will be compared to the totalizer value for the negative flow direction.
The comparison will also be made if the totalizer of the other flow direction is displayed.
18.5.3 Defining the Hysteresis
A hysteresis can be defined for alarm output R1 preventing a constant triggering of the
alarm by measured values fluctuating marginally around the limit.
The hysteresis is a symmetrical range around the limit. The alarm will be activated if the
measured values exceed the upper limit and deactivated if the measured values fall be-
low the lower limit.
example 2:
lower limit = -10 m
3
/h
A measured value of e.g. -11 m
3
/h falls below the limit. The alarm switch-
es.
A measured value of e.g. -9.9 m
3
/h does not fall below the limit. The alarm
does not switch.
Table 18.6: Limit of totalizer
switching
condition
display
comparison
QUANTITY
totalizer
limit
The alarm will switch if the totalizer reaches
the limit.
Note!
The unit of measurement of the limit corresponds to the unit of mea-
surement of the selected physical quantity.
If the unit of measurement of the physical quantity is changed, the
limit has to be converted and entered again.
example 1:
physical quantity volume flow in m
3
/h, quantity limit = 1 m
3
example 2:
physical quantity volume flow in m
3
/h, lower limit = 60 m
3
/h
The unit of measurement of the physical quantity will be changed to m
3
/
min. . The new limit to be entered is 1 m
3
/min.
example:
The limit is 30 m
3
/h and the hysteresis 1 m
3
/h. The alarm will be triggered
for values > 30.5 m
3
/h and deactivated for values < 29.5 m
3
/h.
Enter the hysteresis or 0 (zero) to work without hysteresis.
Press ENTER.
Quantity Limit
1.00
m3
R1 Hysteresis:
1.00
m3/h
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......