10 Basic Measurement
10.7.2 Mounting the Transducers with Mounting Shoes and Tension
Straps
Mounting the Transducers with Clasps
• Cut the tension straps (pipe circumf 120 mm).
• Make sure that part (2) of the clasp is on top of part (1) (see Fig. 10.8). The hooks of
part (2) must be on the outer side of the clasp.
• Pull approx. 2 cm of the tension strap through the slot of the clasp (see Fig. 10.9) to fix
the clasp to the tension strap.
• Bend the end of the tension strap back. Push the other end of the tension strap through
the groove on the top of the mounting shoe.
• Place the clasp on the side of the pipe. Turn the tension strap around the pipe (see Fig.
10.10).
• Place the mounting shoe on the pipe. Hold clasp and mounting shoe with one hand
while pushing the tension strap through parts (2) and (1) of the clasp (see Fig. 10.8).
• Pull the tension strap firmly and engage it in the inner hooks of the clasp.
Fig. 10.8: Clasp
Fig. 10.9: Clasp with tension strap
Note!
The clasp must be in contact completely with the pipe to ensure a
good fixation.
Note!
At pipes with large diameters, use tongs, if necessary, to tense the
tension strap.
1
2
1
2
Summary of Contents for FLUXUS G704
Page 4: ......
Page 235: ...UMG70XV3 4EN 12 01 2009 235 D Certificates D Certificates...
Page 236: ...236 UMG70XV3 4EN 12 01 2009 D Certificates...
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...UMG70XV3 4EN 12 01 2009 D Certificates...
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......