3C
Programming the Generator
CPI Canada Inc.
Page 3C-4
Rev. B Millenia / Indico 100 Series Service Manual Ch # 740904-05
3C.3.1 Setting Time And Date
The SET TIME & DATE menu allows changing or setting of the time and date.
* SET TIME & DATE *
YEAR:
2000
HOUR: 18
MONTH:
9
MIN: 29
DAY:
15
+
-
EXIT
Use these steps to set the time and date.
Step
Action
1.
From the UTILITY menu, select SET TIME & DATE.
2.
Select YEAR and press the + or - buttons to set the year.
3.
Select MONTH and press the + or - buttons to set the month.
4.
Select DAY and press the + or - buttons to set the date.
5.
Select HOUR and press the + or - buttons to set the hour (in 24 hour format).
6.
Select MIN and press the + or - buttons to set the minutes.
7.
Select EXIT to return to the UTILITY menu.
3C.3.2 Error Log
The ERROR LOG menu allows the display of error messages stored in the generator’s error log.
Parameters such as kV, mA, time, receptor, focus, tech selection, field, film screen and fluoro
parameters will be displayed simultaneously on the console LED displays.
* ERROR LOG *
ERROR # 1 OF 18
DATE: 8-19-2000
TIME: 13:09
ERROR CODE: 200
+
ERROR MESSAGE: ANODE HEAT WARN -
EXIT
Use these steps to access the error log.
Step
Action
Result
1.
From the UTILITY menu select ERROR
LOG.
2.
Select ERROR # and press the + or -
buttons to scroll through the error log.
The error code, error message, date and
time of the error will be displayed in the
LCD window, and the associated
parameters will be displayed on the
console displays.
3.
Select EXIT to return to the UTILITY
menu.