4
Acceptance Testing
CPI Canada Inc.
Page 4-12
Rev. F Millenia / Indico 100 Series Service Manual Ch # 740904-08
4.5.1 Generator
Rad
Tests
(Cont)
Step
Action
Result
Check
5.
Repeat step 3 but set the values to
kVp = 125, mA = 200.
kVp = 125 KV
±
3%.
mA = 200 mA
±
4% (Millenia).
mA = 200 mA
±
4% (Indico 100).
Time = 50 ms
±
2 ms.
6.
Select 75 kVp, 200 mA.
Select the exposure times shown below (3
point operation).
(Measure time and mA on the ‘scope and
check that their product is as per the RESULT
column).
Measure time at 75% of the peak kVp
waveform.
A: 10 ms (2 mAs)
B: 20 ms (4 mAs)
C: 63 ms (12 mAs)
D: 100 ms (20 mAs)
A: 2 mAs
±
5%.
B: 4 mAs
±
5%.
C: 12 mAs
±
5%.
D: 20 mAs
±
5%.
7.
Select
75
kVp.
Select the mAs shown below (2 point
operation).
Measure time at 75% of the peak kVp
waveform.
A: 2 mAs (time per LCD display)
B: 8 mAs (time per LCD display)
C: 25 mAs (time per LCD display)
D: 63 mAs (time per LCD display)
NOTE: The time associated with
each mAs setting will vary
depending on generator
configuration.
Use the time displayed in the
LCD window as the reference
for the measurements below.
time per LCD disp
±
(2% + 1ms).
time per LCD disp
±
(2% + 1ms).
time per LCD disp
±
(2% + 1ms).
time per LCD disp
±
(2% + 1ms).
8.
Select 200 mA, 50 ms (3 point operation).
Select the kVp values shown below.
The measurements may be done non-
invasively or from TP6 and TP8 using the
‘scope.
A: 50 kVp
B: 60 kVp
C: 80 kVp
D: 100 kVp
E: 125 kVp
A: 50 kVp
±
(3% + 1 kVp).
B: 60 kVp
±
(3% + 1 kVp).
C: 80 kVp
±
(3% + 1 kVp).
D: 100 kVp
±
(3% + 1 kVp).
E: 125 kVp
±
(3% + 1 kVp).