Programming the Generator
CPI Canada Inc.
Page 3C-14
Rev. B Millenia / Indico 100 Series Service Manual Ch # 740904-05
3C.5.3 Receptor Setup (cont)
•
REM TOMO BUT
Sets the default tomo backup time when tomo is selected via the
REMOTE TOMO SELECT input.
•
SF/LF SWITCH
AUTO: Small or large focus will automatically be selected by the
generator depending on the tube current.
MAN: The operator must manually select small/large focus.
•
AEC BACKUP
Defines the AEC backup mode to be used:
FIXED: The generator will determine the maximum AEC backup
time, not to exceed preset AEC backup mAs/ms values or
system limits. The characters AEC will be displayed in the time
window of the LED display during AEC operation.
MAS: Allows the operator to adjust the AEC backup mAs, not to
exceed preset AEC backup mAs/ms values or system limits.
The mAs value will be displayed in the time window of the LED
display during AEC operation.
MS: Allows the operator to adjust the AEC backup ms, not to
exceed preset AEC backup mAs/ms values or system limits.
The ms value will be displayed in the time window of the LED
display during AEC operation.
•
AEC BACKUP MAS
Sets the maximum backup mAs, to a limit of 500 mAs (600 mAs
for some generator models).
•
AEC BACKUP MS
Sets the maximum back-up ms.
•
DEFAULTS
This selection is available only if MEMORY was set to DEF in
RECEPTOR SETUP menu 2.
THE FOLLOWING SELECTIONS ARE ONLY AVAILABLE IF DEFAULTS WAS ENABLED AS
PREVIOUSLY DESCRIBED.
•
TECHNIQUE
Defines which technique will be defaulted to when a receptor is
selected. Options are AEC, MA, MAS.
•
FOCUS
Defines which focus will be defaulted to when a receptor is
selected. Options are LARGE or SMALL.
•
FILM SCREEN
Defines which film screen will be defaulted to when a receptor is
selected and AEC enabled. Options are film screen 1, 2, or 3.
•
LEFT FIELD
Selects the left field on the AEC device when AEC is selected.
•
CENTER FIELD
As above, but for center field.
•
RIGHT FIELD
As above, but for right field.
•
KV
Selects the default kV value.
•
MA
Selects the default mA value.
•
MS
Selects the default time value.
•
DENSITY
Selects the default density value.