3D
AEC Calibration
CPI Canada Inc.
Page 3D-34
Rev. D Millenia / Indico 100 Series Service Manual Ch # 740904-06
3D.7.0 AEC DENSITY CALIBRATION (Cont)
Step
Action
12.
Set the calibration number for the highest density plus step to the desired relative density
value (example 100, this will result in approximately double the 0 density dose).
13.
Repeat
step
8
and 9.
14.
Select the next lowest density step (i.e. 7) and enter the appropriate calibration number
for that step. Refer to table 3D-15 and notes preceding table 3D-15. Repeat steps 8 and
9.
15.
Repeat the previous step for each remaining density plus step. Use the >> and <<
buttons to navigate through the DENS SETUP menu in order to access all the density
steps.
16.
When the density setup is complete and verified via dose measurements, use the <<
and EXIT buttons to return to the AEC CALIBRATION menu.