CPI Canada Inc.
Installation
2
Indico 100 Series Service Manual Ch # 740895-04 Rev. H
Page 2-33
2.18.2 Initial
Power
Up
1.
Switch on the generator at the console and observe the startup sequence on the console APR
display.
•
MEMORY TEST.... will be displayed
•
HIGH FREQUENCY GENERATOR XX KW WILL BE DISPLAYED (XX will be the kW rating
for that model)
•
The next screen will show console software revision and power software revision
2.
In the generator control cabinet, verify the following:
•
On the generator CPU board verify that the following LEDs are ON (these indicate presence
of the DC rails as indicated):
DS38 +5
V
DS41 +15
V
DS43 -15
V
DS45 +12
V
DS46 -12
V
3.
In the power supply cabinet, verify the following:
•
On the auxiliary board verify that the following LEDs are ON:
CNTCTR CLSD
S.S. OK
+12V
-12V
+/- 35V