2.5.0 LINE
ADJUSTING
TRANSFORMER
An optional line adjusting transformer is available if required to allow 480 VAC generators to operate from
400 VAC mains, or to allow 400 VAC generators to be operated from 480 VAC mains. The line adjusting
transformer is supplied in a separate enclosure. Please consult the factory for further details regarding this
option.
2.6.0 MAJOR
COMPONENT
LAYOUT
Refer to chapter 1E for major component identification and layout.
2.7.0 INSPECTING
THE
HT
TANK
Before continuing with the installation of the generator, the HT oil tank should be inspected as per the
following steps:
•
Verify that there is no obvious shipping damage to the tank (i.e. dents to the tank surface).
•
Carefully check the inside of the generator cabinet and the HT tank for evidence of any oil loss. Refer
to chapter 6 if it is suspected that there has been some loss of oil.
•
Verify that the clamps supporting the HT oil tank are tight.
•
Verify that all the connections to the tank lid are secure.
•
Verify that the rubber vent plug on the top of the tank is snug.
2.8.0 EQUIPMENT
PLACEMENT
2.8.1 Equipment
Cabinet
Place the equipment cabinet in a location that will allow the following:
•
Easy front and side access for service and sufficient clearance at the rear for room interface cables.
Refer to chapter 1C.
•
Air circulation - a minimum height of 1 1/2
″
(35 mm) is recommended to allow airflow underneath the
generator.
•
Stable footing - the leveling feet at the bottom of the cabinet will be used to prevent movement during
normal operation.
•
Close proximity to service disconnect boxes - cables should not be on the floor where they could be
stepped on.
2.8.2 Control
Console
Locate the control console in its intended position and ensure that it is stable. Refer to chapter 1C:
•
If the console is located on a shelf, supply index pins or equivalent hardware to the base of the
console to prevent slipping.
•
Ensure that the console is mounted at a height and angle to allow easy viewing of the displays.
•
If the optional CPI pedestal stand is to be used for the console mounting, follow the mounting
instructions supplied with the stand.
•
Leave sufficient slack in the cabling to the console to allow for future service and maintenance.