Programming the Generator
3C
Millenia / Indico 100 Series Service Manual Ch # 740904-05 Rev. B
Page 3C-13
3C.5.3 Receptor Setup (cont)
Definitions of receptor setup programming as used in this section follow. The selections made
when programming a receptor apply to that receptor only.
•
TUBE
Selects the tube assigned to that receptor. Selecting NONE
disables that receptor.
•
TOMO
Enables or disables tomographic operation (NO is disabled).
•
FLUORO
Enables or disables fluoroscopic operation (NO is disabled).
•
SERIAL
Allows repeated (serial) X-ray exposures without the need to re-
prep after each exposure. Normally used with serial film
changers (NO is disabled).
•
AEC CHANNEL
Defines which AEC channel will be used by the receptor. This
must be set to a valid channel or 0 as noted below. For example
if using an AEC board with only 3 input channels (channels 1 to
3) then selecting AEC channel 4 will cause an error. Selecting 0
disables AEC operation on that receptor.
•
INTERFACE OPTS
Selects pre-defined interface options:
0 = None
1 = InfiMed digital
2 = ATS ESI digital
3 = Gilardoni digital pulsed RAD
4 = Gilardoni digital HCF
5 = Not assigned at this time
•
RECEPTOR SYM
Allows one of six receptor symbols [sym] to be associated with
the selected receptor. Symbol 7 is blank and may be used for
an auxiliary device.
•
FLUORO HANG
Sets the time that the rotor will continue spinning after a fluoro
exposure has terminated.
•
RAD HANG
Sets the time that the rotor will continue spinning after a rad
exposure has terminated.
•
LAST IMAGE HOLD
Sets the time that the exposure will continue after the fluoro
footswitch has been released. This enables a frame store
device to complete the last image.
•
MEMORY
Defines the techniques that will be defaulted to when a receptor
is selected:
YES: The selected receptor will remember it’s last techniques
such that those techniques are displayed when that receptor is
re-selected.
NO: The selected receptor will not remember the last
techniques used on that receptor. The techniques used will be
the same as last used on the previous receptor.
DEF: The techniques used for that receptor will be as
programmed. See receptor setup menu 4 and 5