3E
ABS Calibration
CPI Canada Inc.
Page 3E-14
Rev. D Millenia / Indico 100 Series Service Manual Ch # 740904-07
3E.3.1 Dose Limits (Cont)
8.
Record the kV value as determined in step
7 for the 6.0 mA step.
Max kV limit for 6.0 mA = ______(ABS)
9.
Repeat steps 6 to 8 for 5.0 mA.
Max kV limit for 5.0 mA = ______(ABS)
10.
Repeat steps 6 to 8 for 4.0 mA.
Max kV limit for 4.0 mA = ______(ABS)
11.
Repeat steps 6 to 8 for 3.0 mA.
Max kV limit for 3.0 mA = ______(ABS)
12.
Repeat steps 6 to 8 for 2.0 mA.
Max kV limit for 2.0 mA = ______(ABS)
13.
Repeat steps 6 to 8 for 1.0 mA.
Max kV limit for 1.0 mA = ______(ABS)
The following steps set up the non-ABS kV limits (manual mode).
Step Action
Result
1.
Reselect the 6.0 MA dose limit step in the
LCD display window. Enter the value 6.0
mA in the fluoro control section of the
console.
2.
While observing the dosimeter, make a
fluoroscopy exposure. Adjust the kV via
the fluoro section of the console such that
the maximum permitted dose as recorded
in step 1 (page 13) for non-ABS mode is
not exceeded.
3.
Record the kV value as determined in step
2 for the 6.0 mA step.
Max kV limit for 6.0 mA = ______(MAN)
4.
Repeat steps 1 to 3 for 5.0 mA.
Max kV limit for 5.0 mA = ______(MAN
5.
Repeat steps 1 to 3 for 4.0 mA.
Max kV limit for 4.0 mA = ______(MAN
6.
Repeat steps 1 to 3 for 3.0 mA.
Max kV limit for 3.0 mA = ______(MAN
7.
Repeat steps 1 to 3 for 2.0 mA.
Max kV limit for 2.0 mA = ______(MAN
8.
Repeat steps 1 to 3 for 1.0 mA.
Max kV limit for 1.0 mA = ______(MAN)
9.
Enter the kV limit for non-ABS mode as
recorded in step 3 above (for the 6.0 mA
station) into the 6.0 mA dose limit step in
the LCD display.
Select the 6.0 mA step in the
DOSE LIM - MAN menu, then enter the
required kV value using the + or - buttons
adjacent to the LCD display.
10.
Repeat step 9 for the 5.0 mA to 1.0 mA
stations (non-ABS kV limits).