CPI Canada Inc.
ABS Calibration
3E
Millenia / Indico 100 Series Service Manual Ch # 740904-07 Rev. D
Page 3E-15
3E.3.1 Dose Limits (Cont)
11.
Select << to return to the DOSE LIM-ABS
menu.
12.
Enter the kV limit for ABS mode as
recorded in step 8 (page 14) for the 6.0 mA
station into the 6.0 mA dose limit step in
the the LCD display.
Select the 6.0 mA step in the
DOSE LIM - ABS menu, then enter the
required kV value using the + or - buttons
adjacent to the LCD display.
13.
Repeat step 12 for the 5.0 to 1.0 mA
stations (ABS kV limits)..
14.
Select << to return to the FLUORO SETUP
menu.
3E.3.2 ABS
Defaults
TUBE1 0%HU FLUORO SETUP* MAG:0
DOSE LIMITS
I/I MODES: 2
ABS SETUP
FL-RAD KV XFER: 6
MIN FLUORO KV: 50
+
-
EXIT >> DOSE:0
Step
Action
1.
From the FLUORO SETUP menu select MIN FLUORO KV. Press the + or - buttons
to select the minimum kV to be allowed in fluoro.
2.
Select I/I MODES. Press the + or - buttons to select the number of mag modes in the
I.I. (2 corresponds to 2 mag modes plus normal mode).
3.
Select FL-RAD KV XFER. Press the + or - buttons to select the desired fluoro-rad kV
transfer curve.
This allows selection of one-of-six fluoro to rad kV transfer curves. When in fluoro
operation with ABS on, the fluoro kV value is transferred to the RADIOGRAPHY
section of the console at the end of the fluoro exposure. This presets the rad kV in
preparation for a rapid follow-on radiographic exposure for digital acquisition or spot
film work.
Graphs of the fluoro kV to rad kV transfer function are shown following this step.
Selecting 0 disables this function.