5
Troubleshooting
CPI Canada Inc
Page 5-8 Rev. B Millenia / Indico 100 Series Service Manual Ch # 740904-09
5.2.3 Error
Messages
(Cont)
ERROR CODE
MESSAGE
PROBLEM
ACTION
E058
CONSOLE BUT ERR
Console has detected
exposure exceeded backup
time and terminated
exposure.
Call product support.
E059
HOUSE HEAT LIMIT
X-ray tube housing has
exceeded limit.
Wait for tube to cool.
E060
EXP KV HIGH
KV exceeds high KV
tolerance level.
1. Check the output
of the kV
reference DAC on
the generator
CPU board.
2. Measure the
output of the
generator with a
dynalyzer or a
non-invasive kVp
meter.
E061
EXP KV LOW
KV exceeds low KV
tolerance level.
1. Check the output
of the kV
reference DAC on
the generator
CPU board.
2. Measure the
output of the
generator with a
dynalyzer or a
non-invasive kVp
meter.
E062
EXP_ SW ERROR
The EXP_SW signal on the
generator Interface and
generator CPU board is
enabled when it should be
disabled.
Call product support.
E063
FACTORY DEFAULTS
SW1 switch 8 on the
generator CPU board is set
to default the generator CPU
NVRAM with factory
defaults.
Set SW1 switch 8 to
its non default
position. The
generator will not exit
the initialization
phase until this
switch is set.
E070 SOFTWARE
KEY
ERR
Defective
or
missing
GAL
U29 on generator CPU
board 734573.
Call product support
for new GAL U29.
E100 CAL_MAX
MA
ERR
Maximum
mA
has
been
exceed during auto
calibration.
Repeat auto
calibration and/or
decrease standby
current.