Programming the Generator
3C
Millenia / Indico 100 Series Service Manual Ch # 740904-05 Rev. B
Page 3C-9
3C.5.1 Tube Selection (cont)
NOTE:
BEFORE CHANGING X-RAY TUBE DEFAULT PARAMETERS, PLEASE FILL
IN THE X-RAY TUBE AND GENERATOR PARAMETER WORKSHEET. A
BLANK FORM THAT SHOULD BE PHOTOCOPIED IS LOCATED IN SECTION
3A 3.0. THIS ALLOWS RECORDING OF THE DEFAULT VALUES AND THE
NEW (CHANGED) VALUES.
The next three menus show the default tube limits. These menus appear after a tube has been selected in
the previous steps.
*TUBE 1: RAD60 0.6/1.2 12
º
REV 1.4*
TUBE SPEED: DUAL
MAX SF KW HS: 35.2
MAX SF KW LS: 21.3
MAX LF KW HS: 99.0
MAX LF KW LS: 58.9
+
MAX KV: 150
-
RETURN >>
*TUBE 1: RAD60 0.6/1.2 12
º
REV 1.4*
MAX SF MA: 320
ANODE HU WARNING: 80%
ANODE HU LIMIT: 90%
+
-
<< >>
*TUBE 1: RAD60 0.6/1.2 12
º
REV 1.4*
SF STANDBY: 2.5A
FIL BOOST: 200MS
LF STANDBY: 2.5A
FIL PREHEAT: 800MS
SF MAX: 5.2A
+
LF MAX: 5.5A
-
<< RETURN
Definitions of tube defaults as used in this section.
•
TUBE SPEED
May be altered on dual speed generators. DUAL means that
the generator determines whether to use low speed or high
speed operation.
•
MAX SF KW LS
Sets the maximum small focus low speed kW limit.
•
MAX LF KW LS
Sets the maximum large focus low speed kW limit.
•
MAX KV:
Sets the maximum kV allowed for that tube.
•
MAX SF KW HS
Sets the maximum small focus high speed kW limit.
•
MAX LF KW HS:
Sets the maximum large focus high speed kW limit.