
3E
ABS Calibration
CPI Canada Inc.
Page 3E-22
Rev. D Millenia / Indico 100 Series Service Manual Ch # 740904-07
3E.3.4 Dose1/Dose2 Calibration (Cont)
7.
Reset the original NOMINAL DOSE value
by entering the value recorded in step 1 of
this table into the NOMINAL DOSE
location using the + or - buttons.
8.
Verify each dose (nominal, dose 1, and
dose 2) by initiating fluoro operation and
measuring the I.I. input dose.
9.
Press <<, then EXIT to exit out of
FLUORO SETUP mode. Press EXIT, then
EXIT SETUP to return to the normal
operating mode.
This completes the fluoro calibration.