CPI Canada Inc.
Programming the Generator
3C
Millenia / Indico 100 Series Service Manual Ch # 740904-05 Rev. B
Page 3C-5
3C.3.3 Statistics
The STATISTICS menu shows the tube exposure count, accumulated fluoro hours if applicable, and
the accumulated generator exposure count. This also allows resetting of select counters.
* STATISTICS *
TUBE 1 EXP: 500
RESET TUBE 1 EXP
TUBE 2 EXP: 600
RESET TUBE 2 EXP
FLUORO HOURS: 100
RESET FLUORO HOURS
TOTAL EXP: 1100
EXIT
Use these steps to access the statistics menu. RAD only or 1 tube generators will show a subset of
the above menu items.
Step
Action
1.
From the UTILITY menu select STATISTICS.
2.
To reset the tube exposure counter, select RESET TUBE 1 EXP or RESET TUBE 2
EXP. Select RESET FLUORO HOURS to reset the fluoro exposure counter.
3.
Select EXIT to return to the UTILITY menu.
3C.3.4 Console
The CONSOLE menu allows setting of the items below to suit operator preferences:
* CONSOLE *
SLOW KEY REPEAT: 200MS
LCD SCREEN
MED. KEY REPEAT: 150MS
APR MODE: NO
FAST KEY REPEAT: 75MS
+
SPEAKER VOLUME: 15
-
EXIT
Definitions of console parameters as used in this section.
•
SLOW KEY REPEAT
Determines the speed at which displays change while the
selected key is pressed for the first 5 counts.
•
MED. KEY REPEAT
Determines the speed at which displays change while the
selected key is pressed for the next 5 counts.
•
FAST KEY REPEAT
Determines the speed at which displays change while the
selected key is pressed after 10 counts.
•
SPEAKER VOLUME:
Sets the speaker volume for the control console in the range 1
to 15.
•
LCD SCREEN
Toggles between normal and reverse video for the LCD
display.
•
APR MODE:
NO allows the operator to select an APR view, while still having
the ability to use the TECH SELECT button to manually select
AEC, mAs or mA/ms.
YES only allows the operator to use the preprogrammed
technique selections for APR.