CPI Canada Inc.
Preparing for Installation
1C
Indico 100 Series Service Manual Ch # 740895-02 Rev. D
Page 1C-3
1C.2.4 37.5 kW Three Phase
Line Voltage
400 VAC
±
10%, 3
∅
(for 400 VAC input generators).
480 VAC
±
10%, 3
∅
with optional line adjusting transformer,
(for 400 VAC input generators).
480 VAC
±
10%, 3
∅
(for 480 VAC input generators).
*
SEE NOTE AT THE END OF 1C.2.7
Line Frequency
50/60 Hz.
Momentary Current
95 Amps/phase at 400 VAC.
75 Amps/phase at 480 VAC.
Standby Current
5 Amps.
1C.2.5 50 kW Three Phase
Line Voltage
400 VAC
±
10%, 3
∅
(for 400 VAC input generators).
480 VAC
±
10%, 3
∅
with optional line adjusting transformer,
(for 400 VAC input generators).
480 VAC
±
10%, 3
∅
(for 480 VAC input generators).
*
SEE NOTE AT THE END OF 1C.2.7
Line Frequency
50/60 Hz.
Momentary Current
120 Amps/phase at 400 VAC.
100 Amps/phase at 480 VAC.
Standby Current
5 Amps.
1C.2.6 65 kW Three Phase
Line Voltage
400 VAC
±
10%, 3
∅
(for 400 VAC input generators).
480 VAC
±
10%, 3
∅
with optional line adjusting transformer,
(for 400 VAC input generators).
480 VAC
±
10%, 3
∅
(for 480 VAC input generators).
*
SEE NOTE AT THE END OF 1C.2.7
Line Frequency
50/60 Hz.
Momentary Current
150 Amps/phase at 400 VAC.
120 Amps/phase at 480 VAC.
Standby Current
5 Amps.