CPI Canada Inc.
Programming the Generator
3C
Millenia / Indico 100 Series Service Manual Ch # 740904-05 Rev. B
Page 3C-17
3C.5.4 I/O Configuration (cont)
* OUTPUTS [sym] *
BKY 1 SELECT: --------------- STANDBY STATE
BKY 2 SELECT: ---------------
BKY 3 SELECT: ---------------
TOMO/BKY 4 SEL: ---------------
<<
⇑
>>
* OUTPUTS [sym] *
TOMO/BKY STRT: --------------- STANDBY STATE
ALE: ---------------
COLL. BYPASS: ---------------
<<
⇑
>>
* OUTPUTS [sym] *
ROOM LIGHT: --------------- STANDBY STATE
SPARE: ---------------
<<
⇑
RETURN
The
STATE button on the upper right hand side of the menu selects the current state. The word
STATE will be preceded by the description of the state: for example, STANDBY.
The arrow in the lower middle area points to one of the five states described below. Moving to the
next state is accomplished by pressing the STATE button. The states are as follows:
•
STANDBY
Sets state of the I/O when the generator is in standby or idle
mode. Standby mode also defines the state when the generator
is in fluoroscopic hangover.
•
PREP
Sets state of the I/O when the generator first enters PREP
mode.
•
GEN RDY
Sets state of the I/O when the generator has completed PREP
mode and is ready to expose.
•
RAD EXP
Sets state of the I/O when the generator starts a radiographic
exposure.
•
FLUORO EXP
Sets state of the I/O when the generator starts a fluoroscopic
exposure.