CPI Canada Inc.
System Interfacing
3B
Indico 100 Series Service Manual Ch # 740895-05 Rev. E
Page 3B-3
3B.2.0 LOCATIONS OF INPUTS AND OUTPUTS
3B.2.1 Generator Pictorial Showing Connections
R O O M IN T E R FA C E B O A R D
Figure 3B-1: Generator to room interface
CAUTION:
ENSURE THAT ALL X-RAY TUBE HOUSINGS ARE CONNECTED TO THE GROUND STUD ON
THE HT TANK. USE A SEPARATE GROUND WIRE FOR EACH TUBE, #10 AWG (6 mm
2
) OR
GREATER.