Acceptance Testing
4
Millenia / Indico 100 Series Service Manual Ch # 740904-08 Rev. F
Page 4-15
4.6.0
ACCEPTANCE TESTS (OPTIONAL INTERFACES)
Refer to separate supplements in this manual for further information IF APPLICABLE.
4.7.0
ACCEPTANCE TESTS (AEC)
This section applies only to generators with AEC.
•
Review Section 3D: AEC Calibration
•
Recheck the mAs, Dose, and O.D. as recorded during initial installation. Follow the appropriate steps
in section 3D to verify the AEC calibration.
4.8.0
ACCEPTANCE TESTS (ABS)
This section applies only to non-digital R&F generators with ABS.
•
Review section 3E: ABS calibration.
•
Recheck the dose limits and input dose as described in the procedure.