1C
Preparing for Installation
CPI Canada Inc.
Page 1C-4
Rev. D Indico 100 Series Service Manual Ch # 740895-02
1C.2.7 80 kW Three Phase
Line Voltage
400 VAC
±
10%, 3
∅
(for 400 VAC input generators).
480 VAC
±
10%, 3
∅
with optional line adjusting transformer,
(for 400 VAC input generators).
480 VAC
±
10%, 3
∅
(for 480 VAC input generators).
*
SEE NOTE AT THE END OF 1C.2.7
Line Frequency
50/60 Hz.
Momentary Current
200 Amps/phase at 400 VAC.
180 Amps/phase at 480 VAC.
Standby Current
5 Amps.
*NOTE:
THREE PHASE INDICO 100 GENERATORS ARE AVAILABLE IN 400 VAC AND 480 VAC
MODELS. 400 VAC MODELS MUST BE OPERATED FROM 400 VAC MAINS, OR MAY BE
OPERATED FROM 480 VAC MAINS WITH AN OPTIONAL LINE ADJUSTING
TRANSFORMER. 480 VAC MODELS MUST BE OPERATED FROM 480 VAC MAINS (THESE
ARE THE NOMINAL MAINS VOLTAGES, THE ALLOWED TOLERANCES ARE AS DETAILED
IN THE PREVIOUS TABLES).
1C.2.8 Service Disconnect (All Models)
Refer to the following table for recommended service disconnect ratings.