1A
Introduction
CPI Canada Inc.
Page 1A-4
Rev. E Millenia / Indico 100 Series Service Manual Ch # 740904-00
1A.4.0 APPLICATIONS
RAD
SYSTEMS:
•
Bucky table, table top and off-table radiography
•
Vertical bucky/cassette radiography
•
Conventional tomography
R&F
SYSTEMS
•
Fluoroscopic and spot film applications
•
Tomography with conventional and/or remote R&F tables
•
Optional digital compatible.
1A.5.0 FEATURES
•
High frequency generator.
•
One or two tube operation, Rad or Rad / Fluoro.
•
Single or dual filament supplies.
•
Low speed or dual speed X-ray tube stator supply.
•
Optimal matching of X-ray tubes by PROMs.
•
Repetitive self checks of generator functions, provides display of system faults and operating errors.
•
Optional AEC, up to four inputs.
•
Optional ABS with kVp or kVp/mA fluoro stabilizer.
•
Optional remote fluoro control box for table top use or SFD mounting.
•
X-ray Tube protection. The generator allows setting the following limits:
a) Maximum mA, adjustable for each focal spot.
b) Maximum kVp, adjustable for each X-ray tube.
c) Maximum kW, adjustable for each focal spot.
d) Maximum filament current limit, adjustable for large and small focal spots.
e) Anode heat warning and anode heat alarm levels.
•
Calibration features :
a) Microprocessor design allows all calibration and programming to be performed via the Console.
b) mA
calibration
is
automated.
•
Messages and diagnostic information: For users and service personnel, the generator console
displays various messages indicating status or equipment problems. The user is prompted in case of
errors.
•
Error log stores last 200 errors and associated generator settings.
•
Service and diagnostic information available via a lap-top computer (optional).
•
KVp range:
Radiography 40 to 150 kVp.
Fluoroscopy
40
to
125
kVp.
•
mA range:
Radiography 10 to 320 mA (30 kW), 10 to 400 mA (37.5 kW), 10 to 630 mA (50
kW), 10 TO 800 mA (65 kW) and 10 to 1000 mA (80 kW).
Fluoroscopy
0.5
to
6.0
mA.
•
mAs range:
tube dependent, max 1000 mAs.
•
Time range:
Radiography 2.0 to 6300 ms.
Fluoroscopy 0 to 5 or 0 to 10 minutes.
Refer to product description (end of section 1D) for compatibility and features of this specific generator.