Chapter 3 Repair
69
Maintenance manual
MM-MIXO-EN Ind. 0
04/2012
Mixolab
1.
Remove the left and right casings
2.
Disconnect the equipped pump cable from the CPU board (
A
).
3.
At pump level disconnect the two water tubes linking the pump
to the tank (
B
).
4.
Undo the two attachment nuts (
C
) of the pump and remove it.
Refitting the new equipped pump does not require any particular
instruction.
Notes: - Replacing the small hydraulic tubes is highly recom-
mended.
- A re-calibration of the pump shall be performed, see
Operation 4-7 Dismantling/installing the equipped metering pump
Sheet: 1 / 1
Consumables:
Tools:
Cable ties
-
B
A
C
Summary of Contents for Mixolab
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 6 Maintenance manual MM MIXO EN Ind 0 04 2012...
Page 8: ...8 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 15: ...15 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 2 TROUBLESHOOTING...
Page 16: ...16 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 31: ...31 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 3 REPAIR...
Page 32: ...32 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 38: ...38 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 122: ...122 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 123: ...123 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 4 SPARE PARTS...
Page 124: ...124 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 126: ...126 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 144: ...144 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...