80
Chapter 3 Repair
Maintenance manual
MM-MIXO-EN Ind. 0
04/2012
Mixolab
1.
Remove the left and right casings
2.
Unscrew then remove the torquemeter cable (
A
) at sensor
level.
3.
Cut the cable tie holding the cable on the torquemeter support.
4.
Disconnect the torquemeter cable from the CPU board (
B
).
5.
Extract cable terminals of the torquemeter cable from the con-
nector.
Refitting the new torquemeter cable does not require any particular
instruction.
Operation 6-7 Dismantling/installing torque measurement cable
Sheet: 1 / 1
Consumables:
Tools:
Cable ties
-
B
A