Chapter 3 Repair
71
Maintenance manual
MM-MIXO-EN Ind. 0
04/2012
Mixolab
1.
Remove the left and right casings
, as well as the
central casing
2.
Disconnect the tube from the compressor
(A)
(record previ-
ously its location).
3.
Unplug the compressor connector from the CPU board (
B
).
4.
Undo the attachment screw of the ground cable (
C
) on the
compressor.
5.
Cut the rubber support feet (
D
) from under the compressor
support plate and remove the compressor.
Refitting the new membrane compressor does not require any
particular instruction.
Note: Replacing the tube is recommended.
5 : Pneumatic components
Operation 5-1 Dismantling/installing the membrane compressor
Sheet: 1 / 1
Consumables:
Tools:
Cable ties
-
B
C
A
D
Summary of Contents for Mixolab
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 6 Maintenance manual MM MIXO EN Ind 0 04 2012...
Page 8: ...8 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 15: ...15 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 2 TROUBLESHOOTING...
Page 16: ...16 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 31: ...31 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 3 REPAIR...
Page 32: ...32 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 38: ...38 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 122: ...122 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 123: ...123 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 4 SPARE PARTS...
Page 124: ...124 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 126: ...126 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 144: ...144 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...