Chapter 3 Repair
105
Maintenance manual
MM-MIXO-EN Ind. 0
04/2012
Mixolab
4.
Enter by hand on the right bowl the adjusted calibration rate by
clicking on the corresponding button (see illustration above);
5.
Validate the following confirmation message:
6.
Retransfer the adjusted calibration rate to the mixolab by clic-
king on the purpose double arrow;
7.
Close the "Calibrating the temperature" menu by clicking on
the red cross;
8.
Control and check if the adjusted calibration rate transfer is
done. For doing this, return to part 1-3-2;
9.
Prepare and relauch a test by activating this time the function
"Testing with temperature adjustment" (see part 1).
11 : Adjustments and checks for the Mixolab
Operation 11-4 Calibrating the dough temperature
Sheet: 11/
11
Consumables:
Tools:
-
-
Summary of Contents for Mixolab
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 6 Maintenance manual MM MIXO EN Ind 0 04 2012...
Page 8: ...8 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 15: ...15 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 2 TROUBLESHOOTING...
Page 16: ...16 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 31: ...31 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 3 REPAIR...
Page 32: ...32 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 38: ...38 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 122: ...122 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 123: ...123 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 4 SPARE PARTS...
Page 124: ...124 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 126: ...126 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 144: ...144 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...