ERR 9
"Water metering problem!"
Pump limit stop time-out in filling; limit
stop not reached after15 s
■
Check the pump operation from the
maintenance screen:
- place the nozzle on the tank,
- enter a sufficient value for the amount of
water and press ON
- check that the switch is activated ("ON")
each time the piston is at the back.
■
If the piston does not move, or stops
moving too early, there is a mechanical
failure or a motor failure: check motor wiring
and connections, then replace the pump if
the problem is not solved.
■
If the piston moves normally while the
switch is never activated ("OFF"), there is a
switch failure: check switch wiring and
connections, then replace the pump if the
problem is not solved.
■
If the problem is not solved, replace the
CPU board.
ERR No
Description
Action
See in
Chap.3
Summary of Contents for Mixolab
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 6 Maintenance manual MM MIXO EN Ind 0 04 2012...
Page 8: ...8 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 15: ...15 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 2 TROUBLESHOOTING...
Page 16: ...16 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 31: ...31 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 3 REPAIR...
Page 32: ...32 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 38: ...38 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 122: ...122 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 123: ...123 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 4 SPARE PARTS...
Page 124: ...124 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 126: ...126 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 144: ...144 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...