Chapter 3 Repair
51
Maintenance manual
MM-MIXO-EN Ind. 0
04/2012
Mixolab
1.
Remove the left and right casings
2.
Remove the silicone mastic (
A
) which seals the temperature
sensor to the white plastic plate.
3.
Unplug the "Safety and temperature measurement" connector
(Pt100 temperature sensor) on the CPU board (
B
), then cut
the cable tie.
4.
Remove the temperature sensor from the heating plate (
C
).
5.
Coat the new sensor with silicon heat conductive compound
and insert it into the heating plate.
6.
Put some silicone mastic on the temperature sensor visible
portion at the white plastic plate level.
7.
Re-plug the "Safety and temperature measurement" connector
on the CPU board, then secure the cable a cable tie.
8.
Calibrate the temperature sensor
9.
Reinstall the left and right Mixolab casings
.
Operation 2-2 Dismantling/installing the wired Pt100 temperature sensor
Sheet: 1 / 1
Consumables:
Tools:
Silicone heat conductive compound
Cable ties
A
C
D
Summary of Contents for Mixolab
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 6 Maintenance manual MM MIXO EN Ind 0 04 2012...
Page 8: ...8 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 15: ...15 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 2 TROUBLESHOOTING...
Page 16: ...16 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 31: ...31 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 3 REPAIR...
Page 32: ...32 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 38: ...38 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 122: ...122 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 123: ...123 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 4 SPARE PARTS...
Page 124: ...124 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 126: ...126 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 144: ...144 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...