118
Chapter 3 Repair
Maintenance manual
MM-MIXO-EN Ind. 0
04/2012
Mixolab
First turn off the Mixolab and disconnect the mains supply.
1.
Remove the right casing
.
2.
Disconnect the temperature probe to be tested (see
(A)
)
3.
Measure on the cable’s connector the ohmic resistance of the
probe:
- Bowl block temperature = between terminals [1 or 3] and [2
or 4].
- Tank temperature = between terminals [1 or 4] and [2 or 5].
4.
The resistance should be 100
Ω
at 0°C, with a slope of
0.384
Ω
/°C; i.e. for example for 25°C: 100 + (0.384x25), or
109.6
Ω
.
5.
Once measurements done, re-install the right casing of the
Mixolab
11 : Adjustments and checks for the Mixolab
Operation 11-12 Checking characteristics of a Pt100 temperature probe
Sheet: 1 / 1
Consumables:
Tools:
-
Multimeter
A
Bowl temperature
Tank temperature
CPU
board
Summary of Contents for Mixolab
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 6 Maintenance manual MM MIXO EN Ind 0 04 2012...
Page 8: ...8 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 15: ...15 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 2 TROUBLESHOOTING...
Page 16: ...16 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 31: ...31 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 3 REPAIR...
Page 32: ...32 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 38: ...38 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 122: ...122 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 123: ...123 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 4 SPARE PARTS...
Page 124: ...124 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 126: ...126 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 144: ...144 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...