98
Chapter 3 Repair
Maintenance manual
MM-MIXO-EN Ind. 0
04/2012
Mixolab
■
Testing with temperature adjustment:
❏
Determining the temperature rate:
In this part, we will describe the steps to achieve a calibration rate:
1.
Access the "Configuration" menu, then the "Maintenance"
menu;
2.
In the synoptic view of the "Maintenance" menu, click on "
Tes-
ting without temperature adjustment
» (
A
) (see illustration
below) + enter
1576
: this manipulation allows to activate the
"
Calibrating the temperature
" menu.
3.
Close the maintenance menu, and click on the "
Calibrating
the temperature"
icon (
B
) shown on the illustration below:
11 : Adjustments and checks for the Mixolab
Operation 11-4 Calibrating the dough temperature
Sheet: 4 /
11
Consumables:
Tools:
-
-
A
B
Summary of Contents for Mixolab
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 6 Maintenance manual MM MIXO EN Ind 0 04 2012...
Page 8: ...8 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 15: ...15 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 2 TROUBLESHOOTING...
Page 16: ...16 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 31: ...31 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 3 REPAIR...
Page 32: ...32 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 38: ...38 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 122: ...122 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 123: ...123 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 4 SPARE PARTS...
Page 124: ...124 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 126: ...126 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 144: ...144 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...