Chapter 3 Repair
Maintenance manual
MM-MIXO-EN Ind. 0
04/2012
Mixolab
■
Verifying the calibration
1.
Under the piston icon enter 25ml (the verification will be car-
ried out with 3 samples of 25ml).
2.
Perform the same operations as previously; repeat the calibra-
tion operations from step 9 (calibration for 10ml).
The measurements should be close to 25. The tolerance interval is
24,875 /25,125 ml.
3.
Report the 3 values on the test record and check that they are
within the tolerance values
(K)
.
11 : Adjustments and checks for the Mixolab
Operation 11-9 Calibrating the Mixolab pump
Sheet: 4 / 4
Consumables:
Tools:
-
-
K
Summary of Contents for Mixolab
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 6 Maintenance manual MM MIXO EN Ind 0 04 2012...
Page 8: ...8 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 15: ...15 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 2 TROUBLESHOOTING...
Page 16: ...16 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 31: ...31 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 3 REPAIR...
Page 32: ...32 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 38: ...38 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 122: ...122 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 123: ...123 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 4 SPARE PARTS...
Page 124: ...124 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 126: ...126 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 144: ...144 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...