Chapter 3 Repair
91
Maintenance manual
MM-MIXO-EN Ind. 0
04/2012
Mixolab
1.
Check on the maintenance screen that the bowl temperature
setpoint on the Mixolab is set to 30˚C.
Then let the temperature stabilize for at least 15 minutes.
2.
Place the Plain bowl in the bowl support and lock.
3.
Place the thermometer in the plain bowl into the hole provided
(
A
).
4.
Wait 10 min. then read the temperature on the thermometer.
5.
Compare the thermometer temperature to the bowl tempera-
ture measured by the Mixolab (
B
): they should be the same at
± 0,5°C.
Note: If the temperatures are not the same, go to the maintenance
menu, and proceed as follows.
11 : Adjustments and checks for the Mixolab
Operation 11-2 Calibration the bowl and thermocouple temperatures
Sheet: 1 / 2
Consumables:
Tools:
-
Precision thermometer ±0,1°C
"Plain" bowl
B
A
Summary of Contents for Mixolab
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 6 Maintenance manual MM MIXO EN Ind 0 04 2012...
Page 8: ...8 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 15: ...15 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 2 TROUBLESHOOTING...
Page 16: ...16 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 31: ...31 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 3 REPAIR...
Page 32: ...32 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 38: ...38 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 122: ...122 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 123: ...123 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 4 SPARE PARTS...
Page 124: ...124 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 126: ...126 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 144: ...144 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...