![Chopin Mixolab Maintenance Manual Download Page 76](http://html1.mh-extra.com/html/chopin/mixolab/mixolab_maintenance-manual_2602979076.webp)
76
Chapter 3 Repair
Maintenance manual
MM-MIXO-EN Ind. 0
04/2012
Mixolab
1.
Remove the left and right casings
2.
Unplug the connector of the inductive sensor on the CPU board
(A)
(common connector with the torques measurement cable).
3.
Extract cable terminals of the inductive sensor from the con-
nector.
4.
Hold the sensor rear nut with an open end wrench and loosen
the front nut.
5.
Remove the inductive sensor.
The sensor shall be re-installed so that its front edge is posi-
tioned within the range from
0,8 mm to 1 mm
from the
encoder wheel
(B)
(use of a feeler gauge).
Operation 6-3 Dismantling/installing the inductive sensor
Sheet: 1 / 1
Consumables:
Tools:
Cable ties
Feeler set
A
B
Summary of Contents for Mixolab
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 6 Maintenance manual MM MIXO EN Ind 0 04 2012...
Page 8: ...8 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 15: ...15 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 2 TROUBLESHOOTING...
Page 16: ...16 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 31: ...31 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 3 REPAIR...
Page 32: ...32 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 38: ...38 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 122: ...122 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 123: ...123 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 4 SPARE PARTS...
Page 124: ...124 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 126: ...126 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 144: ...144 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...