20
Chapter 2 Troubleshooting
Maintenance manual
MM-MIXO-EN Ind. 0
04/2012
Mixolab
ERR 16
"Electronic problem on bowl block
heating!"
Permanent bowl block heating
control detected by heating safety
■
Switch the Mixolab off and then on.
■
Problem on CPU board
- Check all connections on the Mixolab CPU
board
- Replace the CPU board if the problem is
not solved.
ERR 21
"Tank temperature too high!"
Cooling time out (after 15 min of
heating absence, tank T˚ > 110% of
setpoint)
■
Confirm on maintenance screen.
■
The lab temperature is too high (it shall be
lower than [setpoint_T – 5°C]) -> modify it
and wait for temperature decrease or
replace the water by colder water.
■
Wrong temperature indication:
- Calibrate the tank temperature
measurement
- Check and replace the sensor if the
problem is not solved.
ERR 22
"Tank temperature too low!"
Heating time out (after 15 min of
heating absence, tank T˚ < 90% of
setpoint)
■
Confirm on maintenance screen.
■
Wrong temperature indication :
- Calibrate the tank temperature
measurement.
- Check and replace the sensor if the
problem is not solved.
■
Insufficient heating :
- Check the heating circuit.
- If necessary, replace the failed resistor.
ERR 23
"Tank empty!"
Icon at the bottom of the screen = red
Fill the Mixolab tank. If there is enough
water:
■
Wrong indication of water level:
- sensor float jammed: remove it, check it
and replace it if necessary.
- Replace the CPU board if the problem is
not solved.
ERR 3
No communication with the Mixolab,
during access to a menu requiring a
dialogue
■
Verify that the Mixolab is correctly
connected and switched on.
■
Check the USB connection (or RS 232
connection) between the PC and the
Mixolab, including internal wiring.
■
Replace the CPU board if the problem is
not solved.
ERR No
Description
Action
See in
Chap.3
Summary of Contents for Mixolab
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 6 Maintenance manual MM MIXO EN Ind 0 04 2012...
Page 8: ...8 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 15: ...15 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 2 TROUBLESHOOTING...
Page 16: ...16 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 31: ...31 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 3 REPAIR...
Page 32: ...32 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 38: ...38 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 122: ...122 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 123: ...123 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 4 SPARE PARTS...
Page 124: ...124 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 126: ...126 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 144: ...144 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...