62
Chapter 3 Repair
Maintenance manual
MM-MIXO-EN Ind. 0
04/2012
Mixolab
1.
Remove the right casing
.
2.
Remove the water injection nozzle (
A
).
3.
Remove the tube ball joint from the central casing (
B
), then
remove the tube.
Note: Record the location of the hydraulic connection on the tank
(see schematic next page)
.
4.
Disconnect the earth terminal from the chassis (
C
).
5.
Unplug all hydraulic connections on the tank.
6.
Cut the cable ties of the electrical cables on the tank.
7.
Unplug the connector of the double level detector on the CPU
board (
D
).
8.
Unplug the connector of the tank heating resistors on the CPU
board (
E
).
9.
Unplug the connector of the tank temperature sensor on the
CPU board (
F
).
4 : Hydraulic components
Operation 4-1 Dismantling/installing the equipped tank
Sheet: 1 / 2
Consumables:
Tools:
Cable ties
-
B
A
C
D
F
E
Summary of Contents for Mixolab
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 6 Maintenance manual MM MIXO EN Ind 0 04 2012...
Page 8: ...8 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 15: ...15 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 2 TROUBLESHOOTING...
Page 16: ...16 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 31: ...31 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 3 REPAIR...
Page 32: ...32 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 38: ...38 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 122: ...122 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 123: ...123 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 4 SPARE PARTS...
Page 124: ...124 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 126: ...126 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Page 144: ...144 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...