un

®

TEMPERATURE MANAGEMENT SYSTEM

MODEL 5000

CALIBRATION TEST UNIT (CTU)

OPERATOR’S MANUAL 

Summary of Contents for Arctic Sun 5000

Page 1: ...Arctic Sun TEMPERATURE MANAGEMENT SYSTEM MODEL 5000 CALIBRATION TEST UNIT CTU OPERATOR S MANUAL...

Page 2: ......

Page 3: ...3 2 Basic CTU Operation 6 3 3 Controls 7 3 4 Connections 7 3 5 Understanding the Test and Calibration Sequence 7 3 5 1 Calibration Factor Longevity 7 3 5 2 Initiation 7 3 5 3 Pre Warm and Flow Check...

Page 4: ...rature Management System to enable it to test and calibrate itself The sole active function of the CTU is to provide an external heat source to raise the circulating water temperature for troubleshoot...

Page 5: ...tomatic Stage 2 Patient Temperature Calibration 5 minutes operator assistance required Stage 3 Water Temperature Calibration 18 minutes operator assistance required at completion Stage 4 Water Tempera...

Page 6: ...luid delivery lines and pads The CTU can be used to semi automatically test and calibrate the Control Module s two patient temperature measurement inputs the temperature out function and the internal...

Page 7: ...The three electrical connections are labeled PT1 PT2 and TO o PT1 corresponds to the primary patient temperature receptacle on the control module o PT2 corresponds to the secondary patient temperatur...

Page 8: ...us setpoints while PT1 measures the temperature of the water flowing through the CTU This reference temperature is then used to check the system s internal water temperature sensors The heater element...

Page 9: ...Maintenance The Calibration and Test Unit contains no operator maintainable parts Other than periodic cleaning of the exterior surfaces no maintenance is required Cleaning instructions Spray a mild d...

Page 10: ...C T U M A N U A L 10 TEMPERATURE MANAGEMENT SYSTEM APPENDIX A Specifications Mains Input 115VAC 60Hz 4 5A 230VAC 50 60Hz 2 25A 100VAC 50 60Hz 5 0A Fuse 6 3A Slow Blow 2 Power 500W...

Page 11: ...re 1 off of conversion chart table at 1 C Error 44 Water temperature 1 off of conversion table at 47 C Error 50 Water temperature 2 is out of range at 6 C Error 51 Water temperature 2 is out of range...

Page 12: ...12 TEMPERATURE MANAGEMENT SYSTEM GUIDE DU DISPOSITIF DE TEST D TALONNAGE...

Page 13: ...2 Fonctionnement de base du CTU 17 3 3 Commandes 17 3 4 Branchements 17 3 5 Comprendre la s quence de test et d talonnage 17 3 5 1 Long vit du facteur d talonnage 17 3 5 2 D marrage 17 3 5 3 V rifica...

Page 14: ...our lui permettre de proc der lui m me aux tests et l talonnage La seule fonction active du CTU est de fournir des fins de d pannage et sous la direction du personnel de l assistance technique Medivan...

Page 15: ...lonnage de la pression d entr e 12 minutes automatique Stade 2 talonnage de la temp rature du patient 5 minutes aide de l op rateur requise Stade 3 talonnage de la temp rature de l eau 18 minutes aide...

Page 16: ...dispositif qui tient dans la main et se branche directement sur le port de la ligne de distribution de fluide Il remplace les lignes de distribution de fluide et les pads du mod le 5000 Le CTU peut tr...

Page 17: ...fant S4 est un interrupteur de rappel normalement ouvert qui alimente les deux l ments chauffants en courant alternatif de 250 W lorsqu il est enfonc et maintenu en position MARCHE Le c ble d alimenta...

Page 18: ...l ordre de 1 0 C La diff rence entre la moyenne de PT1 et la temp rature de sortie moyenne telle que mesur e par le connecteur PT2 est de l ordre de 0 8 C La temp rature de sortie telle que mesur e pa...

Page 19: ...C 0 3 C 40 C 0 3 C Temp rature de l eau T3 surveillance entr e 3 v rifications 0 5 C 6 C 0 6 C 0 5 C 28 C 0 5 C 0 5 C 40 C 0 5 C Tableau 3 1 Donn es de v rification du mod le 5000 Chapitre 4 Maintena...

Page 20: ...EMENT SYSTEM ANNEXE A Caract ristiques techniques Entr e secteur 115 VCA 60 Hz 4 5 A 230 VCA 50 60 Hz 2 25 A 100 VCA 50 60 Hz 5 0 A Fusible 6 3 A fusion lente 2 Alimentation 500 W GUIDE DU DISPOSITIF...

Page 21: ...re de l eau 1 hors du tableau de conversion 47 C Erreur 50 La temp rature de l eau 2 se situe hors limites 6 C Erreur 51 La temp rature de l eau 2 se situe hors limites 28 C Erreur 52 La temp rature d...

Page 22: ...22 TEMPERATURE MANAGEMENT SYSTEM WARTUNGSHANDBUCH F R DAS KALIBRIERUNGSPR FGER T CTU...

Page 23: ...2 Grundlegende Bedienung des Kalibrierungspr fger ts CTU 27 3 3 Bedienelemente 27 3 4 Anschl sse 27 3 5 Erl uterung des Test und Kalibrierablaufs 27 3 5 1 Lebensdauer der Kalibrierfaktoren 27 3 5 2 E...

Page 24: ...um dieses zur Selbsttestung und kalibrierung zu bef higen Die einzige aktive Funktion des CTU besteht in der Bereitstellung einer externen W rmequelle um die Temperatur des zirkulierenden Wasser zu er...

Page 25: ...brierung 5 Minuten Eingreifen des Anwenders erforderlich Phase 3 Wassertemperatur Kalibrierung 18 Minuten Eingreifen des Anwenders erforderlich nach Abschluss Phase 4 Wassertemperatur Kalibrierung Hei...

Page 26: ...trument das zur Kalibrierung und Funktionspr fung des Steuerungsmoduls des Arctic Sun Temperaturmanagementsystems Modell 5000 dient Das handtellergro e Ger t wird direkt mit dem Fl ssigkeitszufuhrleit...

Page 27: ...em Patiententemperatur Kreislauf Position B verbindet einen Parallelwiderstand mit dem internen Thermistor wodurch die Temperatur bei der niedrigen Wassertemperatureinstellung hochgeschaltet wird und...

Page 28: ...ngstemperatur misst Ausgangstemperatur ist ein Echo der Patiententemperatur 1 Dieser Test erfordert offenkundig eine pr zise Messung durch die Patiententemperatur Kan le Daher m ssen die Patiententemp...

Page 29: ...t 3 Pr fungen 0 5 C 6 C 0 5 C 0 5 C 28 C 0 5 C 0 3 C 40 C 0 3 C Wassertemperatur T3 Monitor Ein 3 Pr fungen 0 5 C 6 C 0 6 C 0 5 C 28 C 0 5 C 0 5 C 40 C 0 5 C Tabelle 3 1 Eichdaten Modell 5000 Abschnit...

Page 30: ...EM ANHANG A Technische Daten Netzstrom 115 V Wechselstrom 60 Hz 4 5 A 230 V Wechselstrom 50 60 Hz 2 25 A 100 V Wechselstrom 50 60 Hz 5 0 A Sicherung 6 3 A tr ge 2 Leistung 500 W WARTUNGSHANDBUCH F R D...

Page 31: ...r 43 Wassertemperatur 1 bei 1 C von Umrechnungstabelle abweichend Fehler 44 Wassertemperatur 1 bei 47 C von Umrechnungstabelle abweichend Fehler 50 Wassertemperatur 2 bei 6 C au erhalb des zul ssigen...

Page 32: ...32 TEMPERATURE MANAGEMENT SYSTEM MANUALE UTC...

Page 33: ...i base dell UTC 36 3 3 Comandi 37 3 4 Collegamenti 37 3 5 Comprensione della sequenza di test e calibrazione 37 3 5 1 Durata del fattore di calibrazione 37 3 5 2 Inizializzazione 37 3 5 3 Pre riscalda...

Page 34: ...tic Sun affinch questo possa condurre procedure di test e calibrazione su se stesso L unica funzionalit attiva ricoperta dall UTC fornire una fonte di calore esterna per aumentare la temperatura dell...

Page 35: ...one della temperatura del paziente 5 minuti assistenza dell operatore necessaria Fase 3 calibrazione della temperatura dell acqua 18 minuti assistenza dell operatore necessaria al completamento Fase 4...

Page 36: ...ega direttamente alla porta della linea di erogazione dei fluidi sostituendosi alle linee di erogazione dei fluidi e ai cuscinetti del modello 5000 L UTC pu essere usata per eseguire test e calibrazio...

Page 37: ...peratura acqua bassa in modo che il circuito di temperatura del paziente sia in grado di rilevarla L interruttore di accensione del dispositivo di riscaldamento S4 un interruttore a impulsi generalmen...

Page 38: ...ale a 0 01 C Se una verifica qualsiasi non soddisfa i suddetti criteri entro 75 secondi viene visualizzato un messaggio di errore 3 5 6 Verifica della temperatura dell acqua Il sistema richiede all ut...

Page 39: ...raccomandata L unit test e calibrazione non contiene parti che possono essere sottoposte a manutenzione da parte dell operatore Non richiesto alcun intervento di manutenzione in aggiunta alla pulizia...

Page 40: ...TURE MANAGEMENT SYSTEM APPENDICE A Specifiche Ingresso dell alimentazione di rete 115 V CA 60 Hz 4 5 A 230 V CA 50 60 Hz 2 25 A 100 V CA 50 60 Hz 5 0 A Fusibile 6 3 A ritardato 2 Alimentazione 500 W M...

Page 41: ...e 44 La temperatura dell acqua 1 non rientra nella tabella di conversione a 47 C Errore 50 La temperatura dell acqua 2 fuori dall intervallo a 6 C Errore 51 La temperatura dell acqua 2 fuori dall inte...

Page 42: ...42 TEMPERATURE MANAGEMENT SYSTEM MANUAL DE LA UPC...

Page 43: ...3 Controles 47 3 4 Conexiones 47 3 5 Descripci n de la secuencia de prueba y calibraci n 47 3 5 1 Duraci n del factor de calibraci n 47 3 5 2 Inicio 47 3 5 3 Precalentamiento y comprobaci n del flujo...

Page 44: ...orma que este pueda comprobarse y calibrarse La nica funci n activa de la UPC es ofrecer una fuente de calor externa que aumente la temperatura del agua circulante para la resoluci n de problemas bajo...

Page 45: ...utom tico Etapa 2 calibraci n de la temperatura del paciente 5 minutos se requiere asistencia del operario Etapa 3 calibraci n de la temperatura del agua 18 minutos se requiere asistencia del operario...

Page 46: ...rata de un aparato del tama o de una mano que se conecta directamente al puerto del tubo de administraci n de l quido en lugar de los tubos de administraci n de l quido y las almohadillas del Modelo 5...

Page 47: ...ra del paciente El interruptor de encendido de los calefactores S4 es un interruptor temporal normalmente abierto que aplica corriente alterna CA a los dos calefactores de 250 W cuando se mantiene pul...

Page 48: ...atura media de salida seg n la medici n de PT2 y el valor respectivo de temperatura de la UPC est dentro de 1 0 C La diferencia entre la PT1 media y la temperatura media de salida seg n la medici n de...

Page 49: ...a T2 monitor salida 3 comprobaciones 0 5 C 6 C 0 5 C 0 5 C 28 C 0 5 C 0 3 C 40 C 0 3 C Temperatura del agua T3 monitor entrada 3 comprobaciones 0 5 C 6 C 0 6 C 0 5 C 28 C 0 5 C 0 5 C 40 C 0 5 C Tabla...

Page 50: ...EMPERATURE MANAGEMENT SYSTEM MANUAL DE LA UPC AP NDICE A Especificaciones Entrada el ctrica 115 V CA 60 Hz 4 5 A 230 V CA 50 60 Hz 2 25 A 100 V CA 50 60 Hz 5 0 A Fusible 6 3 A con retardo 2 Potencia 5...

Page 51: ...Error 43 Temperatura del agua 1 fuera de la tabla de conversi n a 1 C Error 44 Temperatura del agua 1 fuera de la tabla de conversi n a 47 C Error 50 Temperatura del agua 2 fuera del intervalo a 6 C E...

Page 52: ...52 TEMPERATURE MANAGEMENT SYSTEM C T U H A N D L E I D I N G...

Page 53: ...nleiding 56 3 2 Basiswerking CTU 56 3 3 Bedieningselementen 57 3 4 Aansluitingen 57 3 5 Inzicht in de test en kalibratieprocedure 57 3 5 1 Gebruiksduur kalibratiefactor 57 3 5 2 Start 57 3 5 3 Voorver...

Page 54: ...perature Management System zodat het systeem zichzelf kan testen en kalibreren De enige actieve functie van de CTU is te voorzien in een externe warmtebron om de temperatuur van het circulerende water...

Page 55: ...kalibratie ingangsdruk 12 minuten automatisch Fase 2 Kalibratie pati nttemperatuur 5 minuten assistentie gebruiker vereist Fase 3 Kalibratie watertemperatuur 18 minuten assistentie gebruiker vereist...

Page 56: ...ature Management System Model 5000 Het is een unit van handformaat die rechtstreeks op de vloeistofafgiftelijnpoort wordt aangesloten waarbij de unit de plaats inneemt van de Model 5000 vloeistofafgif...

Page 57: ...selstroom doorgeeft naar de twee 250 W verwarmers zodra de schakelaar in de AAN positie ingedrukt wordt gehouden De wisselstroomkabel moet op de CTU en een geschikte wandcontactdoos zijn aangesloten o...

Page 58: ...er dan 0 8 C De Temperatuur uitgaand gemeten bij PT2 is stabiel binnen een marge van 0 01 C Als een controle niet binnen 75 seconden aan deze criteria voldoet verschijnt er een foutmelding 3 5 6 Contr...

Page 59: ...en testunit bevatten geen onderdelen die door de gebruiker kunnen worden onderhouden Behalve het periodiek reinigen van de buitenkant is er geen onderhoud nodig Reinigingsinstructies Spray een milde...

Page 60: ...ERATURE MANAGEMENT SYSTEM C T U H A N D L E I D I N G BIJLAGE A Specificaties Netstroom 115 VAC 60 Hz 4 5 A 230 VAC 50 60 Hz 2 25 A 100 VAC 50 60 Hz 5 0 A Zekering 6 3 A langzaam doorbrandend 2 Stroom...

Page 61: ...abel van 1 C Fout 44 Watertemperatuur 1 valt buiten de conversiegrafiektabel van 47 C Fout 50 Watertemperatuur 2 is buiten bereik van 6 C Fout 51 Watertemperatuur 2 is buiten bereik van 28 C Fout 52 W...

Page 62: ...62 TEMPERATURE MANAGEMENT SYSTEM M A N U A L D A U T C...

Page 63: ...sico da UTC 66 3 3 Controlos 67 3 4 Liga es 67 3 5 Compreender a sequ ncia de teste e de calibra o 67 3 5 1 Longevidade do fator da calibra o 67 3 5 2 Arranque 67 3 5 3 Aquecimento pr vio e verifica o...

Page 64: ...o sistema realizar um autoteste e uma autocalibra o A nica fun o ativa da UTC fornecer uma fonte de calor externa para elevar a temperatura da gua circulante para fins de resolu o de problemas sob a o...

Page 65: ...erifica o do fluxo e Calibra o da press o de entrada 12 minutos autom tico Etapa 2 Calibra o da temperatura do paciente 5 minutos assist ncia do operador necess ria Etapa 3 Calibra o da temperatura da...

Page 66: ...Sun Modelo 5000 Trata se de uma unidade port til que ligada diretamente porta da linha de distribui o de fluido assumindo o lugar as linhas de distribui o de fluido e almofadas do Modelo 5000 A UTC p...

Page 67: ...mperatura do paciente O interruptor de alimenta o do aquecedor S4 um interruptor tempor rio normalmente aberto que aplica corrente CA aos dois aquecedores de 250 W quando premindo e mantido na posi o...

Page 68: ...o os seguintes crit rios s o cumpridos A diferen a entre a m dia da Sa da da Temperatura conforme medido pela PT2 e o respetivo valor da temperatura da UTC estiver dentro de 1 0 C A diferen a entre a...

Page 69: ...rifica es 0 5 C 6 C 0 5 C 0 5 C 28 C 0 5 C 0 3 C 40 C 0 3 C Temperatura da gua T2 Monitor Sa da 3 verifica es 0 5 C 6 C 0 5 C 0 5 C 28 C 0 5 C 0 3 C 40 C 0 3 C Temperatura da gua T3 Monitor Entrada 3...

Page 70: ...TEMPERATURE MANAGEMENT SYSTEM M A N U A L D A U T C ANEXO A Especifica es Entrada el trica 115 VCA 60 Hz 4 5 A 230 VCA 50 60 Hz 2 25 A 100 VCA 50 60 Hz 5 0 A Fus vel 6 3 A Queima lenta 2 Pot ncia 500...

Page 71: ...a tabela do gr fico de convers o a 1 C Erro 44 Temperatura da gua 1 ausente da tabela do gr fico de convers o a 47 C Erro 50 Temperatura da gua 2 fora do intervalo a 6 C Erro 51 Temperatura da gua 2 f...

Page 72: ...72 TEMPERATURE MANAGEMENT SYSTEM M A N U A L D A U T C...

Page 73: ...b sico da UTC 76 3 3 Controles 77 3 4 Conex es 77 3 5 Compreender a sequ ncia de teste e de calibra o 77 3 5 1 Longevidade do fator da calibra o 77 3 5 2 Inicializa o 77 3 5 3 Aquecimento pr vio e ve...

Page 74: ...a permitir ao sistema realizar um autoteste e uma autocalibra o A nica fun o ativa da UTC fornecer uma fonte de calor externa para elevar a temperatura da gua circulante para fins de resolu o de probl...

Page 75: ...o do fluxo e Calibra o da press o de entrada 12 minutos autom tico Etapa 2 Calibra o da temperatura do paciente 5 minutos assist ncia do operador necess ria Etapa 3 Calibra o da temperatura da gua 18...

Page 76: ...c Sun Modelo 5000 Trata se de uma unidade port til que ligada diretamente porta da linha de distribui o de fluido assumindo o lugar das linhas de distribui o de fluido e almofadas do Modelo 5000 A UTC...

Page 77: ...nte O interruptor de alimenta o do aquecedor S4 um interruptor tempor rio normalmente aberto que aplica corrente CA aos dois aquecedores de 250 W quando pressionado e mantido na posi o LIGADO O cabo d...

Page 78: ...peratura do paciente Assim as verifica es da temperatura do paciente devem primeiro ser conclu das com xito Cada verifica o conclu da quando os seguintes crit rios s o cumpridos A diferen a entre a m...

Page 79: ...do Paciente Sa da 4 verifica es 0 2 C 10 C 0 2 C 0 2 C 25 C 0 2 C 0 2 C 35 C 0 2 C 0 3 C 40 C 0 3 C Temperatura da gua T1 Controle Sa da 3 verifica es 0 5 C 6 C 0 5 C 0 5 C 28 C 0 5 C 0 3 C 40 C 0 3 C...

Page 80: ...TEMPERATURE MANAGEMENT SYSTEM M A N U A L D A U T C ANEXO A Especifica es Entrada el trica 115 VAC 60 Hz 4 5 A 230 VAC 50 60 Hz 2 25 A 100 VAC 50 60 Hz 5 0 A Fus vel 6 3 A Queima lenta 2 Pot ncia 500...

Page 81: ...gr fico de convers o a 1 C Erro 44 Temperatura da gua 1 ausente da tabela do gr fico de convers o a 47 C Erro 50 Temperatura da gua 2 fora da faixa a 6 C Erro 51 Temperatura da gua 2 fora da faixa a...

Page 82: ...82 TEMPERATURE MANAGEMENT SYSTEM C T U...

Page 83: ...4 1 1 CTU 84 1 2 84 1 3 84 1 4 85 1 5 85 2 85 2 1 85 2 2 85 3 86 3 1 86 3 2 CTU 86 3 3 87 3 4 87 3 5 87 3 5 1 87 3 5 2 87 3 5 3 87 3 5 4 88 3 5 5 88 3 5 6 88 3 5 7 88 3 5 8 89 4 89 4 1 89 4 2 89 4 3 8...

Page 84: ...NAGEMENT SYSTEM C T U 1 1 1 CTU Arctic Sun 5000 CTU CTU CTU Arctic Sun 5000 Arctic Sun CTU 1 2 CTU Arctic Sun CTU Medivance 1 3 CTU 1 1 CTU CTU Arctic Sun ETL ETL Intertek C22 2 61010 1 UL 61010 1 50...

Page 85: ...o 5 95 o 5 95 o 375 mmHg 795 mm Hg o 375 mmHg 795 mm Hg IPX0 1 5 CTU 2 2 1 90 1 12 2 5 3 18 4 33 5 25 2 2 A CTU B PT1 5000 1 PT2 5000 2 TO 5000 TEMP OUT E Arctic Sun 5000 H Arctic Sun 5000 A H CTU CT...

Page 86: ...86 TEMPERATURE MANAGEMENT SYSTEM C T U 2 4 Arctic Sun 5000 2 5 Arctic Sun 5000 CTU 2 6 CTU 1 2 7 CTU 2 3 2 8 CTU 3 3 1 CTU Arctic Sun 5000 5000 CTU 3 2 CTU CTU CTU...

Page 87: ...TU 5000 CTU 3 3 S1 S2 S3 S4 CTU S1 A B 10 8 C 25 5 C 34 5 C 40 0 C CTU S2 A 1 2 1 CTU 2 1 S3 A B S4 250 W ON CTU 3 4 PT1 PT2 TO o PT1 o PT2 o TO PT1 PT2 PT1 PT2 PT1 PT2 TO 3 5 Arctic Sun 5000 3 5 1 3...

Page 88: ...PSI 0 5 PSI 3 5 4 1 PT1 1 CTU CTU 0 8 C 0 01 C 60 1 CTU 4 CTU CTU 0 14 C CTU 2 PT2 S2 3 5 5 1 CTU PT1 PT2 1 PT2 CTU 1 0 C PT1 PT2 0 8 C PT2 0 01 C 75 3 5 6 S3 PT1 CTU 15 1 6 C 2 28 C 3 40 5 C 4 38 C 1...

Page 89: ...C 0 3 C 2 2 PT2 4 0 3 C 10 C 0 3 C 0 3 C 25 C 0 3 C 0 1 C 35 C 0 1 C 0 3 C 40 C 0 3 C 4 0 2 C 10 C 0 2 C 0 2 C 25 C 0 2 C 0 2 C 35 C 0 2 C 0 3 C 40 C 0 3 C T1 3 0 5 C 6 C 0 5 C 0 5 C 28 C 0 5 C 0 3 C...

Page 90: ...90 TEMPERATURE MANAGEMENT SYSTEM C T U A 115 VAC 60 Hz 4 5 A 230 VAC 50 60 Hz 2 25 A 100 VAC 50 60 Hz 5 0 A 6 3 A 2 500 W...

Page 91: ...25 C 23 2 35 C 24 2 40 C 25 2 26 2 30 2 31 32 40 1 6 C 41 1 28 C 42 1 40 5 C 43 1 1 C 44 1 47 C 50 2 6 C 51 2 28 C 52 2 40 5 C 53 2 1 C 54 2 47 C 60 3 6 C 61 3 28 C 62 3 40 5 C 63 3 1 C 64 3 47 C 70...

Page 92: ...92 TEMPERATURE MANAGEMENT SYSTEM C T U V E J L E D N I N G...

Page 93: ...betjening 86 3 3 Betjeningsknapper 87 3 4 Tilslutninger 87 3 5 Forst else af test og kalibreringssekvensen 87 3 5 1 Kalibreringsfaktorens holdbarhed 88 3 5 2 Initiering 88 3 5 3 Forvarmning og kontrol...

Page 94: ...rolsystemet s det kan teste og kalibrere sig selv CTU ens eneste aktive funktion er at tilvejebringe en ekstern varmekilde til at h ve temperaturen af det cirkulerende vand med henblik p fejlfinding u...

Page 95: ...g kalibrering af indl bstryk 12 minutter automatisk Fase 2 Kalibrering af patienttemperatur 5 minutter brugerassistance p kr vet Fase 3 Kalibrering af vandtemperatur 18 minutter brugerassistance p kr...

Page 96: ...det for model 5000 s v sketilf rselsslanger og puder CTU en kan anvendes til halvautomatisk test og kalibrering af kontrolmodulets to m leindgange til patienttemperatur udl bstemperaturfunktionen og m...

Page 97: ...iske tilslutninger er m rket PT1 PT2 og TO o PT1 svarer til det prim re patienttemperaturstik p kontrolmodulet o PT2 svarer til det sekund re patienttemperaturstik o TO svarer til udl bstemperaturstik...

Page 98: ...forskellige indstillingspunkter mens PT1 m ler temperaturen p det vand der l ber gennem CTU en Denne referencetemperatur bruges efterf lgende til kontrol af systemets interne vandtemperatursensorer Va...

Page 99: ...Anbefalet vedligeholdelse Kalibrerings og testenheden indeholder ingen dele som kan vedligeholdes af brugeren Vedligeholdelse er ikke p kr vet ud over j vnlig reng ring udvendigt Reng ringsanvisninge...

Page 100: ...TEMPERATURE MANAGEMENT SYSTEM C T U V E J L E D N I N G APPENDIKS A Specifikationer Indgangsstr m 115 V AC 60 Hz 4 5 A 230 V AC 50 60 Hz 2 25 A 100 V AC 50 60 Hz 5 0 A Sikring 6 3 A tr g 2 Effekt 500...

Page 101: ...Vandtemperatur 1 uden for konverteringsdiagramtabel ved 1 C Fejl 44 Vandtemperatur 1 uden for konverteringstabel ved 47 C Fejl 50 Vandtemperatur 2 er uden for omr de ved 6 C Fejl 51 Vandtemperatur 2 e...

Page 102: ...102 TEMPERATURE MANAGEMENT SYSTEM C T U H A N D B O K...

Page 103: ...l ggande handhavande av CTU 106 3 3 Reglage 107 3 4 Anslutningar 107 3 5 F rst test och kalibreringssekvensen 107 3 5 1 Kalibreringsfaktorns livsl ngd 107 3 5 2 Initiering 107 3 5 3 F rv rmning och fl...

Page 104: ...tem s att det kan testa och kalibrera sig sj lvt Den enda aktiva funktionen f r CTU n r att utg ra en extern v rmek lla f r att h ja temperaturen p det cirkulerande vattnet f r fels kning under v gled...

Page 105: ...nuter automatiskt Skede 2 Kalibrering av patienttemperatur 5 minuter anv ndarens assistans kr vs Skede 3 Kalibrering av vattentemperatur 18 minuter anv ndarens assistans kr vs vid slutf randet Skede 4...

Page 106: ...v tsketillf rselslangarna och dynorna p modell 5000 CTU n kan anv ndas f r halvautomatisk testning och kalibrering av kontrollmodulens b da ing ngar f r patienttemperaturm tning funktionen utg ende t...

Page 107: ...m ste vara t t utan l ckage f r ett korrekt test De tre elektriska anslutningarna r m rkta PT1 PT2 och TO o PT1 motsvarar det prim ra patienttemperaturuttaget p kontrollmodulen o PT2 motsvarar den sek...

Page 108: ...s i manuellt l ge under hela detta test N r systemet k rs kontrolleras vattentemperaturen vid olika brytpunkter samtidigt som PT1 m ter temperaturen p det vatten som fl dar genom CTU n Denna referenst...

Page 109: ...0 Sektion 4 Underh ll 4 1 Rekommenderat underh ll Kalibrerings och testenheten inneh ller inte n gra delar som kan underh llas av operat ren Ut ver regelbunden reng ring av de yttre ytorna kr vs inget...

Page 110: ...110 TEMPERATURE MANAGEMENT SYSTEM C T U H A N D B O K BILAGA A Specifikationer Str mf rs rjning 115 V AC 60 Hz 4 5 A 230 V AC 50 60Hz 2 25 A 100 V AC 50 60Hz 5 0 A S kring 6 3A tr g 2 Effekt 500 W...

Page 111: ...ringsdiagramtabell vid 1 C Fel 44 Vattentemperatur 1 utanf r konverteringsdiagramtabell vid 47 C Fel 50 Vattentemperatur 2 utanf r omr de vid 6 C Fel 51 Vattentemperatur 2 utanf r omr de vid 28 C Fel...

Page 112: ...112 TEMPERATURE MANAGEMENT SYSTEM KALIBROINTITESTILAITTEEN K YTT OPAS...

Page 113: ...erusk ytt 116 3 3 Ohjaimet 117 3 4 Liit nn t 117 3 5 Testi ja kalibrointisekvenssin selitys 117 3 5 1 Kalibrointitekij n kestoik 117 3 5 2 Aloittaminen 117 3 5 3 Esil mmitys ja virtauksen tarkistus 11...

Page 114: ...autteen joiden avulla j rjestelm testaa ja kalibroi itsens CTU n ainoa aktiivinen toiminto on toimia ulkoisena l mm nl hteen kiertoveden l mp tilan nostamiseksi vianm rityst varten Medivancen teknisen...

Page 115: ...Potilaan l mp tilan kalibrointi 5 minuuttia k ytt j n toimenpiteit tarvitaan Vaihe 3 Veden l mp tilan kalibrointi 18 minuuttia k ytt j n toimenpiteit tarvitaan loppuvaiheessa Vaihe 4 Veden l mp tilan...

Page 116: ...malliin 5000 nesteensy tt letkut ja kiinnittimet CTU ta voidaan k ytt puoliautomaattisesti konrollointimoduulin kahden potilaan l mp tilamittauksen sis ntulon l mp tilan ulostulotoiminnon ja sis isen...

Page 117: ...kinn t ovat PT1 PT2 ja TO o PT1 vastaa ensisijaista potilaan l mp tila anturin vastaanotinta kontrollointimoduulissa o PT2 vastaa toissijaista potilaan l mp tila anturin liitint o TO vastaa l mp tilan...

Page 118: ...m n k ydess veden l mp tilaa ohjataan eri asetuspisteiss PT1 n mitatessa CTU n l pi virtaavan veden l mp tilaa T m n viitel mp tilan avulla tarkistetaan sitten j rjestelm n sis iset vesil mp tila antu...

Page 119: ...0 vahvistustiedot Osa 4 Kunnossapito 4 1 Suositellut kunnossapitotoimet Kalibrointi ja testilaitteessa ei ole k ytt j n kunnossapidett vi osia Muita kunnossapitotoimia kuin s nn llinen ulkopintojen pu...

Page 120: ...RATURE MANAGEMENT SYSTEM KALIBROINTITESTILAITTEEN K YTT OPAS LIITE A Tekniset tiedot Verkkovirtasy tt 115 V AC 60 Hz 4 5 A 230 V AC 50 60 Hz 2 25 A 100 V AC 50 60 Hz 5 0 A Sulake 6 3 A Slow Blow 2 Vir...

Page 121: ...en l mp tila 1 ei ole muuntotaulukon alueella 1 C ssa Virhe 44 Veden l mp tila 1 ei ole muuntotaulukon alueella 47 C ssa Virhe 50 Veden l mp tila 2 ei ole alueella 6 C ssa Virhe 51 Veden l mp tila 2 e...

Page 122: ...122 TEMPERATURE MANAGEMENT SYSTEM C T U H N D B O K...

Page 123: ...Grunnleggende CTU drift 126 3 3 Kontroller 127 3 4 Tilkoblinger 127 3 5 Forst test og kalibreringssekvensen 127 3 5 1 Lang levetid p kalibreringsfaktor 127 3 5 2 Start 127 3 5 3 Forh ndsvarming og fl...

Page 124: ...Sun temperaturh ndteringssystemet for gj re det i stand til teste og kalibrere seg selv Den eneste aktive funksjonen til CTU er gi en ekstern varmekilde for ke sirkulasjonsvanntemperaturen for feils...

Page 125: ...automatisk Trinn 2 Patienttemperaturkalibrering 5 minutter operat rassistanse n dvendig Trinn 3 Vanntemperaturkalibrering 18 minutter operat rassistanse n dvendig ved fullf ring Trinn 4 Vanntemperatu...

Page 126: ...r enhet som kobles direkte til v skeleveringsslangeporten som tar plassen til modell 5000 v skeleveringsslanger og pads CTU kan brukes til halvautomatisk test og kalibrering av kontrollmodulens to pas...

Page 127: ...en m v re tett uten lekkasjer for sikre en egnet test De tre elektriske tilkoblingene er merket PT1 PT2 og TO o PT1 tilsvarer prim r pasienttemperaturkontakt p kontrollmodulen o PT2 tilsvarer sekund r...

Page 128: ...ved ulike settpunkter mens PT1 m ler temperaturen til vannet som str mmer gjennom CTU Denne referansetemperaturen brukes deretter til kontrollere systemets interne vanntemperatursensorer Varmeelement...

Page 129: ...1 Anbefalt vedlikehold Kalibrerings og testenheten inneholder ingen deler som kan vedlikeholdes av operat ren Bortsett fra rutinemessig rengj ring av de utvendige overflatene er ikke noe annet vedlik...

Page 130: ...130 TEMPERATURE MANAGEMENT SYSTEM C T U H N D B O K VEDLEGG A Spesifikasjoner Str minngang 115 VAC 60 Hz 4 5 A 230 VAC 50 60 Hz 2 25 A 100 VAC 50 60 Hz 5 0 A Sikring 6 3 A treg sikring 2 Effekt 500 W...

Page 131: ...gsoversikttabell ved 1 C Feil 44 Vanntemperatur 1 utenfor konverteringstabell ved 47 C Feil 50 Vanntemperatur 2 er utenfor omr det ved 6 C Feil 51 Vanntemperatur 2 er utenfor omr det ved 28 C Feil 52...

Page 132: ...132 TEMPERATURE MANAGEMENT SYSTEM P O D R C Z N I K O B S U G I U T K...

Page 133: ...uga UTK 136 3 3 Elementy sterowania 137 3 4 Po czenia 137 3 5 Obja nienie sekwencji testowania i kalibracji 137 3 5 1 Okres wa no ci wsp czynnika kalibracji 137 3 5 2 Inicjalizacja 137 3 5 3 Nagrzewan...

Page 134: ...Sun aby umo liwi mu samodzielne wykonanie testu i kalibracji Jedyn aktywn funkcj urz dzenia jest kierowanie do zewn trznego r d a ciep a dania podniesienia temperatury kr cej wody w celu rozwi zania p...

Page 135: ...acji trwa ok 90 minut Dzia anie operatora jest niezb dne jedynie na niekt rych etapach procesu Etap 1 nagrzewanie kontrola przep ywu kalibracja ci nienia wlotu 12 minut automatyczny Etap 2 kalibracja...

Page 136: ...c Sun model 5000 Urz dzenie stanowi jednostk wielko ci d oni Pod czane jest ono bezpo rednio do portu linii dostarczaj cej p yn zamiast linii dostarczaj cej p yn oraz wk adek modelu 5000 UTK mo na wyk...

Page 137: ...a by o mo liwe przew d zasilaj cy pr dem przemiennym musi by pod czony do UTK oraz do odpowiedniego gniazda zasilania 3 4 Po czenia Nale y wykona pod czenie hydrauliczne i trzy pod czenia elektryczne...

Page 138: ...Je eli kt rakolwiek kontrola nie spe ni tych kryteri w w ci gu 75 sekund zostanie wy wietlony komunikat o b dzie 3 5 6 Kontrola temperatury wody System monituje u ytkownika o zmian pozycji S3 na pocz...

Page 139: ...ci kt re mog by poddawane konserwacji przez operatora Nie s wymagane inne czynno ci konserwacyjne poza okresowym czyszczeniem powierzchni zewn trznych Instrukcja czyszczenia rozpyli roztw r agodnego d...

Page 140: ...TURE MANAGEMENT SYSTEM P O D R C Z N I K O B S U G I U T K ZA CZNIK A Dane techniczne Zasilanie 115 V AC 60 Hz 4 5 A 230 V AC 50 60 Hz 2 25 A 100 V AC 50 60 Hz 5 0 A Bezpiecznik 6 3 A bezw adno ciowy...

Page 141: ...poza zakresem przy 40 5 C B d 43 Temperatura wody 1 poza tabel konwersji przy 1 C B d 44 Temperatura wody 1 poza tabel konwersji przy 47 C B d 50 Temperatura wody 2 poza zakresem przy 6 C B d 51 Temp...

Page 142: ...142 TEMPERATURE MANAGEMENT SYSTEM C T U K Z I K N Y V...

Page 143: ...szint kezel se 146 3 3 Vez rl k 147 3 4 Csatlakoz sok 147 3 5 A tesztel si s kalibr l si szakaszok megismer se 147 3 5 1 Kalibr ci s t nyez lettartam 147 3 5 2 Kezdeti l p sek 147 3 5 3 El meleg t s s...

Page 144: ...r sz m ra lehet v t ve hogy az tesztelje s kalibr lja nmag t A CTU egyetlen akt v funkci ja az hogy a Medivance m szaki t mogat szem lyzet ir ny t s val k ls h forr st biztos tson a kering v zh m rs k...

Page 145: ...s 2 szakasz Betegh m rs klet kalibr l s 5 perc kezel i fel gyelet sz ks ges 3 szakasz V zh m rs klet kalibr l s 18 perc kezel i fel gyelet sz ks ges befejez skor 4 szakasz V zh m rs klet kalibr l s me...

Page 146: ...z kapcsol dik tv ve az 5000 es t pus k sz l k folyad ksz ll t vezet keinek s bet teinek hely t A CTU f lautomata m dban teszteli s kalibr lja a vez rl egys g k t betegh m rs klet m r si bemenet t a ki...

Page 147: ...l 3 4 Csatlakoz sok Egy hidraulikus s a h rom elektromos kapcsolatot kell kialak tani a vez rl egys ggel hogy a k sz l k megfelel en m k dj n a vez rl egys g szoftverfunkci ival A hidraulikus kapcsola...

Page 148: ...3 5 6 A v zh m rs klet ellen rz se A rendszer figyelmezteti a felhaszn l t hogy m dos tsa az S3 helyzet t az alacsony h m rs kleti alap rt k be ll t s nak kezdet n majd ll tsa azt vissza az alacsony h...

Page 149: ...m tartalmaz a kezel ltal karbantarthat alkatr szeket A k ls fel letek rendszeres tiszt t s n k v l semmilyen karbantart s nem sz ks ges A tiszt t ssal kapcsolatos utas t sok Permetezzen enyhe mos szer...

Page 150: ...RE MANAGEMENT SYSTEM C T U K Z I K N Y V A F GGEL K M szaki adatok H l zati bemenet 115 VAC 60 Hz 4 5 A 230 VAC 50 60 Hz 2 25 A 100 VAC 50 60 Hz 5 0 A Biztos t k 6 3 A k sleltetett biztos t k 2 Teljes...

Page 151: ...zat rt keit 44 hiba Az 1 v zh m rs klet 47 C on t ll pi a konverzi s t bl zat rt keit 50 hiba Az 2 v zh m rs klet 6 C on a tartom nyon k v l van 51 hiba Az 2 v zh m rs klet 28 C on a tartom nyon k v l...

Page 152: ...152 TEMPERATURE MANAGEMENT SYSTEM P R U K A K C T U...

Page 153: ...dn provoz CTU 156 3 3 Ovlada e 157 3 4 P ipojen 157 3 5 Porozum n sekvenci test a kalibrace 157 3 5 1 Doba pou itelnosti kalibra n ch parametr 157 3 5 2 Spu t n 157 3 5 3 P edeh t a kontrola pr toku 1...

Page 154: ...by se syst m samo inn testoval a kalibroval Jedinou aktivn funkc CTU je to e p edstavuje extern zdroj tepla kter zvy uje teplotu cirkuluj c vody p i e en probl m pod veden m person lu technick podpory...

Page 155: ...a 5 minut zapot eb asistence obsluhy St dium 3 Kalibrace teploty vody 18 minut zapot eb asistence obsluhy p i dokon en St dium 4 Kalibrace teploty vody kontrola oh va e 33 minut zapot eb asistence obs...

Page 156: ...zauj m m sto plnic ch hadi ek a pol t k modelu 5000 CTU lze pou vat k semiautomatick mu testov n a kalibraci dvou m ic ch vstup teploty pacienta funkce teplota venku a intern ch m ic ch kan l teploty...

Page 157: ...n z suvce teploty pacienta na dic m modulu o PT2 odpov d sekund rn mu konektoru teploty pacienta o TO odpov d konektoru teploty venku na dic m modulu TO m jin elektrick konektor ne PT1 a PT2 a proto j...

Page 158: ...b syst m v ru n m re imu Zat mco je syst m spu t n teplota vody je regulov na tak aby dos hla n kolika c lov ch hodnot zat mco PT1 m teplotu vody prot kaj c skrz CTU Tato referen n teplota se pot pou...

Page 159: ...ifika n data pro model 5000 st 4 dr ba 4 1 Doporu en dr ba Kalibra n a testovac jednotka neobsahuje dn sti kter m e udr ovat obsluha Krom pravideln ho i t n zevn ch povrch nen zapot eb dn dr ba Pokyny...

Page 160: ...160 TEMPERATURE MANAGEMENT SYSTEM P R U K A K C T U P LOHA A Specifikace Vstupn s ov nap jen 115 VAC 60 Hz 4 5 A 230 VAC 50 60 Hz 2 25 A 100 VAC 50 60 Hz 5 0 A Pojistka 6 3 A pomal 2 P kon 500 W...

Page 161: ...a 43 Teplota vody 1 mimo p evodn tabulku p i 1 C Chyba 44 Teplota vody 1 mimo p evodn tabulku p i 47 C Chyba 50 Teplota vody 2 je mimo rozsah p i 6 C Chyba 51 Teplota vody 2 je mimo rozsah p i 28 C Ch...

Page 162: ...162 TEMPERATURE MANAGEMENT SYSTEM C T U K I L A V U Z U...

Page 163: ...risi 166 3 1 Giri 166 3 2 Temel CTU Operasyonu 166 3 3 Kontroller 167 3 4 Ba lant lar 167 3 5 Test ve Kalibrasyon Sekans n Anlama 167 3 5 1 Kalibrasyon Fakt r Uzun mr 167 3 5 2 Ba latma 167 3 5 3 n Is...

Page 164: ...k Y netim Sistemine uyar m ve geribildirim sa lamaktad r CTU nun tek aktif i levi Medivance teknik destek personelinin k lavuzlu unda sorun giderme amac yla devridaim su s cakl n y kseltmek i in bir...

Page 165: ...rasyonu 12 dakika otomatik 2 A ama Hasta S cakl Kalibrasyonu 5 dakika operat r yard m gereklidir 3 A ama Su S cakl Kalibrasyonu 18 dakika operat r yard m gereklidir tamamlama s ras nda 4 A ama Su S ca...

Page 166: ...rol Mod l n n iki hasta s cakl k l m giri ini s cakl k k i levini ve dahili su s cakl l m kanallar n yar otomatik olarak test ve kalibre etmekte kullan labilir Ayr ca Kontrol Mod l n n ak h z ve giri...

Page 167: ...asta s cakl prizine denk d er o PT2 ikinci hasta s cakl konekt r ne denk d er o TO kontrol mod l ndeki s cakl k k konekt r ne denk d er TO nun PT1 ve PT2 den farkl bir u emas vard r ve bu nedenle bunl...

Page 168: ...tli ayar noktalar nda kontrol edilir ve bu s rada PT1 CTU dan akan suyun s cakl n l er Bu referans s cakl k daha sonra sistemin dahili su s cakl k sens rlerini kontrol etmekte kullan l r Is tma ve so...

Page 169: ...on ve Test nitesi operat r n bak m yapabilece i bir par a i ermez D y zeylerin periyodik temizli i d nda bir bak m yap lmas gerekmez Temizlik talimatlar Yumu ak bir deterjan sol syonunu bir havluya p...

Page 170: ...170 TEMPERATURE MANAGEMENT SYSTEM C T U K I L A V U Z U EK A Spesifikasyonlar ebeke Giri i 115VAC 60Hz 4 5A 230VAC 50 60Hz 2 25A 100VAC 50 60Hz 5 0A Sigorta 6 3A Yava S nen 2 G 500W...

Page 171: ...rme tablosunu d nda Hata 44 Su s cakl 1 47 C de d n t rme tablosunu d nda Hata 50 Su s cakl 2 6 C de aral k d nda Hata 51 Su s cakl 2 28 C de aral k d nda Hata 52 Su s cakl 2 40 5 C de aral k d nda Ha...

Page 172: ...172 TEMPERATURE MANAGEMENT SYSTEM M A N U A L C T U...

Page 173: ...e 166 3 2 Func ionarea de baz a CTU 166 3 3 Comenzi 167 3 4 Racorduri 167 3 5 n elegerea secven ei de testare i calibrare 167 3 5 1 Longevitatea factorilor de calibrare 167 3 5 2 Ini iere 167 3 5 3 Ve...

Page 174: ...i permite acestuia s efectueze propria testare i calibrare Singura func ie activ a CTU este s asigure o surs de c ldur extern pentru a cre te temperatura apei circulante pentru depanare sub ndrumarea...

Page 175: ...faze Faza 1 nc lzire Verificare flux i Calibrare presiune de admisie 12 minute automat Faza 2 Calibrare temperatur pacient 5 minute este necesar asisten a operatorului Faza 3 Calibrare temperatur ap...

Page 176: ...strare fluid i al padelelor pentru Modelul 5000 CTU poate fi utilizat pentru testarea i calibrarea semiautomat a celor dou intr ri de m surare a temperaturii pacientului a func iei temperatur de ie ir...

Page 177: ...scurgeri pentru a asigura testarea adecvat Cele trei racorduri electrice sunt etichetate PT1 PT2 i TO o PT1 corespunde buc ei principale de temperatur a pacientului de pe modulul de control o PT2 core...

Page 178: ...n rii sistemului temperatura apei este controlat la diferite valori de referin n timp ce PT1 m soar temperatura apei care curge prin CTU Aceast temperatur de referin este utilizat apoi pentru a verifi...

Page 179: ...brare i testare nu con ine componente care pot fi ntre inute de c tre operator Nu sunt necesare alte proceduri de mentenan n afar de cur area periodic a suprafe elor exterioare Instruc iuni de cur are...

Page 180: ...RATURE MANAGEMENT SYSTEM M A N U A L C T U ANEXA A Specifica ii Putere absorbit din re ea 115 V c a 60 Hz 4 5 A 230 V c a 50 60 Hz 2 25 A 100 V c a 50 60 Hz 5 0 A Siguran 6 3 A cu aprindere lent 2 Put...

Page 181: ...roarea 42 Temperatur 1 ap n afara intervalului la 40 5 C Eroarea 43 Temperatur 1 ap n afara tabelului de conversie la 1 C Eroarea 44 Temperatur 1 ap n afara tabelului de conversie la 47 C Eroarea 50 T...

Page 182: ...182 TEMPERATURE MANAGEMENT SYSTEM P R R U K A P R E J E D N O T K U C T U...

Page 183: ...prev dzky jednotky CTU 186 3 3 Ovl dacie prvky 187 3 4 Pripojenia 187 3 5 Vysvetlenie testovacej a kalibra nej sekvencie 187 3 5 1 Trv cnos kalibra n ho faktora 187 3 5 2 Inici cia 187 3 5 3 Predhria...

Page 184: ...ykon va vlastn testovanie a kalibr ciu Jedin akt vna funkcia jednotky CTU spo va v plnen funkcie extern ho tepeln ho zdroja na zv enie teploty cirkuluj cej vody na ely rie enia probl mov pod veden m p...

Page 185: ...tomaticky F za 2 Kalibr cia teploty pacienta 5 min t vy aduje si s innos zo strany oper tora F za 3 Kalibr cia teploty vody 18 min t vy aduje si s innos zo strany oper tora po dokon en F za 4 Kalibr c...

Page 186: ...eploty Arctic Sun model 5000 Ide o zariadenie ktor sa zmest do ruky a ktor sa prip ja priamo k portu rozvodu na pr vod kvapaliny pri om nahr dza pr vodn rozvody na pr vod kvapaliny a podlo ky modelu 5...

Page 187: ...vala v s innosti so softv rov mi funkciami kontroln ho modulu Hydraulick pr pojka je peci lne vyroben diel ktor sa prip ja priamo ku kontroln mu modulu Tesnenie mus dolieha a nesmie vykazova niky aby...

Page 188: ...e v manu lnom re ime Po as prev dzky syst mu sa teplota vody reguluje v r znych bodoch nastavenia zatia o PT1 meria teplotu vody pretekaj cej cez jednotku CTU T to referen n teplota sa n sledne pou ij...

Page 189: ...obsahuje iadne s asti ktor by oper tor mohol svojpomocne podrobova dr be Okrem pravideln ho istenia vonkaj ch povrchov sa nevy aduje iadne dr ba Postup istenia Nastriekajte jemn roztok istiaceho prost...

Page 190: ...MANAGEMENT SYSTEM P R R U K A P R E J E D N O T K U C T U PR LOHA A pecifik cie Nap jac vstup 115 V AC 60 Hz 4 5 A 230 V AC 50 60 Hz 2 25 A 100 V AC 50 60 Hz 5 0 A Poistka 6 3 A s pomal m vypa ovan m...

Page 191: ...40 5 C Chyba 43 Teplota vody 1 mimo tabu ky konverznej sch my pri 1 C Chyba 44 Teplota vody 1 mimo konverznej tabu ky pri 47 C Chyba 50 Teplota vody 2 je mimo rozsah pri 6 C Chyba 51 Teplota vody 2 j...

Page 192: ...192 TEMPERATURE MANAGEMENT SYSTEM...

Page 193: ...4 1 2 194 1 3 194 1 4 195 1 5 195 2 195 2 1 195 2 2 195 3 196 3 1 196 3 2 196 3 3 197 3 4 197 3 5 197 3 5 1 197 3 5 2 198 3 5 3 198 3 5 4 198 3 5 5 198 3 5 6 198 3 5 7 199 3 5 8 199 4 200 4 1 200 4 2...

Page 194: ...194 TEMPERATURE MANAGEMENT SYSTEM 1 1 1 Arctic Sun 5000 Arctic Sun 5000 Arctic Sun 1 2 Arctic Sun Medivance 1 3 1 1 Arctic Sun ETL ETL Intertek 22 2 61010 1 UL 61010 1 50 C 30 C 95 05...

Page 195: ...F o 30 C 50 C 20 F 120 F o 5 95 o 5 95 o 375 795 o 375 795 IPX0 1 5 2 2 1 90 1 12 2 5 3 18 4 33 5 25 2 2 A B 5000 PT1 1 C 5000 PT2 2 D 5000 TO TEMP OUT E Arctic Sun 5000 F G H Arctic Sun 5000 I A H J...

Page 196: ...196 TEMPERATURE MANAGEMENT SYSTEM 2 3 Arctic Sun 5000 2 4 Arctic Sun 5000 2 5 Arctic Sun 5000 2 6 1 2 7 2 3 2 8 3 3 1 Arctic Sun 5000 5000 3 2...

Page 197: ...197 R u s s i a n 1 5000 3 3 S1 S2 S3 S4 S1 A B 10 8 C 25 5 C D 34 5 C 40 0 C F S2 A 1 2 1 2 1 S3 B S4 250 3 4 PT1 PT2 TO o PT1 o PT2 o TO TO PT1 PT2 PT1 PT2 PT1 PT2 TO 3 5 Arctic Sun 5000 3 5 1...

Page 198: ...4 3 5 3 10 28 C 55 10 500 600 1 0 7 0 12 0 01 0 5 3 5 4 1 PT1 1 0 8 C 0 01 C 60 1 4 D 0 14 C 2 PT2 S2 3 5 5 1 PT1 PT2 1 PT2 1 0 C PT1 PT2 0 8 C PT2 0 01 C 75 3 5 6 S3 PT1 15 1 6 C 2 28 C 3 40 5 C 4 38...

Page 199: ...3 C 10 C 0 3 C 0 3 C 25 C 0 3 C 0 1 C 35 C 0 1 C 0 3 C 40 C 0 3 C 2 2 PT2 4 0 3 C 10 C 0 3 C 0 3 C 25 C 0 3 C 0 1 C 35 C 0 1 C 0 3 C 40 C 0 3 C 4 0 2 C 10 C 0 2 C 0 2 C 25 C 0 2 C 0 2 C 35 C 0 2 C 0...

Page 200: ...200 TEMPERATURE MANAGEMENT SYSTEM 4 4 1 4 2 4 3 800 526 4455 5 5 8 30 18 00 877 267 2314 866 840 9776 Arctic Sun 4 4 WEEE Bard A 115 60 4 5 A 230 50 60 2 25 A 100 50 60 5 0 A 6 3 A 2 500...

Page 201: ...2 2 25 C 23 2 35 C 24 2 40 C 25 2 26 2 30 2 31 32 40 1 6 C 41 1 28 C 42 1 40 5 C 43 1 1 C 44 1 47 C 50 2 6 C 51 2 28 C 52 2 40 5 C 53 2 1 C 54 2 47 C 60 3 6 C 61 3 28 C 62 3 40 5 C 63 3 1 C 64 3 47 C...

Page 202: ...202 TEMPERATURE MANAGEMENT SYSTEM C T U...

Page 203: ...4 1 2 204 1 3 204 1 4 205 1 5 205 2 205 2 1 205 2 2 205 3 206 3 1 206 3 2 CTU 206 3 3 207 3 4 207 3 5 207 3 5 1 207 3 5 2 207 3 5 3 207 3 5 4 207 3 5 5 208 3 5 6 208 3 5 7 208 3 5 8 208 4 209 4 1 209...

Page 204: ...EMENT SYSTEM C T U 1 1 1 CTU CTU Arctic Sun 5000 CTU CTU CTU Arctic Sun 5000 Arctic Sun CTU 1 2 CTU Arctic Sun CTU Medivance 1 3 CTU 1 1 CTU ETL Arctic Sun CTU C22 2 No 61010 1 UL 61010 1 ETL Intertek...

Page 205: ...o 5 95 o 5 95 o 375mmHg 795mmHg o 375mmHg 795mmHg IPX0 1 5 CTU 2 2 1 90 1 12 2 5 3 18 4 33 5 25 2 2 A CTU B PT1 5000 1 C PT2 5000 2 D TO TEMP OUT 5000 E Arctic Sun 5000 E G H Arctic Sun 5000 I CTU CT...

Page 206: ...206 TEMPERATURE MANAGEMENT SYSTEM C T U 2 4 Arctic Sun 5000 2 5 Arctic Sun 5000 CTU 2 6 CTU 1 2 7 CTU 2 3 2 8 CTU 3 3 1 CTU Arctic Sun 5000 5000 CTU 2 3 2 CTU CTU CTU CTU...

Page 207: ...TU 2 C 1 S3 A B S4 AC 2 250W AC CTU 3 4 1 3 3 PT1 PT2 TO o PT1 o PT2 o TO TO PT1 PT2 PT1 PT2 PT1 PT2 PT1 PT2 PT1 PT2 TO 3 5 Arctic Sun 5000 3 5 1 3 5 2 1 10 C 12 C 2 10 C 4 CTU 1 3 5 3 10 28 C 55 10 5...

Page 208: ...B 25 54 C C 34 52 C D 39 98 C E 0 00 C F 0 00 C G 0 00 C H 2900 3 5 8 3 1 300ml 300ml 500ml 0 3psi P 0 3psi 1 1 PT1 4 0 3 C 10 C 0 3 C 0 3 C 25 C 0 3 C 0 1 C 35 C 0 1 C 0 3 C 40 C 0 3 C 2 2 PT2 4 0 3...

Page 209: ...9 J A P A N E S E 4 4 1 CTU 4 2 CTU 2 CTU 4 3 800 526 4455 5 5 8 30am 6 00pm 877 267 2314 24 866 840 9776 Arctic Sun 4 4 WEEE Bard A 115VAC 60Hz 4 5A 230VAC 50 60Hz 2 25A 100VAC 50 60Hz 5 0A 6 3A 2 50...

Page 210: ...21 2 10 C 22 2 25 C 23 2 35 C 24 2 40 C 25 2 26 2 30 2 31 32 40 1 6 C 41 1 28 C 42 1 40 5 C 43 1 1 C 44 1 47 C 50 2 6 C 51 2 28 C 52 2 40 5 C 53 2 1 C 54 2 47 C 60 3 6 C 61 3 28 C 62 3 40 5 C 63 3 1...

Page 211: ...211 J A P A N E S E...

Page 212: ...212 TEMPERATURE MANAGEMENT SYSTEM C T U...

Page 213: ...1 CTU 214 1 2 214 1 3 215 1 4 215 1 5 215 2 215 2 1 215 2 2 215 3 216 3 1 216 3 2 CTU 216 3 3 217 3 4 217 3 5 217 3 5 1 217 3 5 2 217 3 5 3 217 3 5 4 217 3 5 5 217 3 5 6 218 3 5 7 218 3 5 8 218 4 219...

Page 214: ...ANAGEMENT SYSTEM C T U 1 1 1 CTU Arctic Sun 5000 CTU CTU CTU Arctic Sun 5000 Arctic Sun CTU 1 2 CTU Arctic Sun CTU Medivance 1 3 CTU 1 1 CTU ETL Arctic Sun CTU ETL Intertek C22 2 61010 1 UL 61010 1 50...

Page 215: ...F 120 F o 5 95 o 5 95 o 375 mmHg 795 mmHg o 375 mmHg 795 mmHg IPX0 1 5 CTU 2 2 1 90 1 12 2 5 3 18 4 33 5 25 2 2 A CTU B PT1 5000 1 C PT2 5000 2 D TO TEMP OUT 5000 E Arctic Sun 5000 F G H Arctic Sun 50...

Page 216: ...216 TEMPERATURE MANAGEMENT SYSTEM C T U 2 4 Arctic Sun 5000 2 5 CTU Arctic Sun 5000 2 6 CTU 1 2 7 CTU 2 3 2 8 CTU 3 3 1 CTU Arctic Sun 5000 5000 CTU 3 2 CTU CTU CTU CTU CTU 1 CTU...

Page 217: ...250 CTU 3 4 PT1 PT2 TO o PT1 o PT2 o TO TO PT1 PT2 PT1 PT2 PT1 PT2 TO 3 5 Arctic Sun 5000 3 5 1 3 5 2 1 10 C 12 C 2 10 C 4 CTU 3 5 3 10 28 C 55 10 500ml min CTU H 600 ml min 7 0 psi 1 0 psi 12 0 01 PS...

Page 218: ...00 C G 0 00 C H 2900 3 5 8 3 1 300 ml min 300 ml min 500 ml min 0 3 psi P 0 3 psi 1 1 PT1 4 0 3 C 10 C 0 3 C 0 3 C 25 C 0 3 C 0 1 C 35 C 0 1 C 0 3 C 40 C 0 3 C 2 2 PT2 4 0 3 C 10 C 0 3 C 0 3 C 25 C 0...

Page 219: ...E S E S I M P L I F I E D 4 4 1 CTU 4 2 CTU CTU 4 3 800 526 4455 5 5 8 30 6 00 877 267 2314 24 866 840 9776 Arctic Sun 4 4 WEEE Bard A 115 VAC 60 Hz 4 5A 230 VAC 50 60 Hz 2 25A 100 VAC 50 60 Hz 5 0A...

Page 220: ...21 2 10 C 22 2 25 C 23 2 35 C 24 2 40 C 25 2 26 2 30 2 31 32 40 1 6 C 41 1 28 C 42 1 40 5 C 43 1 1 C 44 1 47 C 50 2 6 C 51 2 28 C 52 2 40 5 C 53 2 1 C 54 2 47 C 60 3 6 C 61 3 28 C 62 3 40 5 C 63 3 1...

Page 221: ...221 C H I N E S E S I M P L I F I E D...

Page 222: ...222 TEMPERATURE MANAGEMENT SYSTEM C T U...

Page 223: ...1 1 CTU 224 1 2 224 1 3 224 1 4 225 1 5 225 2 225 2 1 225 2 2 225 3 226 3 1 226 3 2 CTU 226 3 3 227 3 4 227 3 5 227 3 5 1 227 3 5 2 227 3 5 3 227 3 5 4 227 3 5 5 227 3 5 6 228 3 5 7 228 3 5 8 228 4 22...

Page 224: ...ANAGEMENT SYSTEM C T U 1 1 1 CTU Arctic Sun 5000 CTU CTU CTU Arctic Sun 5000 Arctic Sun CTU 1 2 CTU Arctic Sun CTU Medivance 1 3 CTU 1 1 CTU ETL Arctic Sun CTU ETL Intertek C22 2 61010 1 UL 61010 1 50...

Page 225: ...F 120 F o 5 95 o 5 95 o 375 mmHg 795 mmHg o 375 mmHg 795 mmHg IPX0 1 5 CTU 2 2 1 90 1 12 2 5 3 18 4 33 5 25 2 2 A CTU B PT1 5000 1 C PT2 5000 2 D TO TEMP OUT 5000 E Arctic Sun 5000 F G H Arctic Sun 5...

Page 226: ...226 TEMPERATURE MANAGEMENT SYSTEM C T U 2 4 Arctic Sun 5000 2 5 CTU Arctic Sun 5000 2 6 CTU 1 2 7 CTU 2 3 2 8 CTU 3 3 1 CTU Arctic Sun 5000 5000 CTU 3 2 CTU CTU CTU CTU CTU 1 CTU CTU 5000...

Page 227: ...C 250W AC CTU 3 4 PT1 PT2 TO o PT1 o PT2 o TO TO PT1 PT2 PT1 PT2 PT1 PT2 TO 3 5 Arctic Sun 5000 3 5 1 3 5 2 1 10 C 12 C 2 10 C 4 CTU 3 5 3 10 28 C 55 10 500ml CTU H 600ml 1 0 psi 7 0 psi 12 0 01 PSI 0...

Page 228: ...C F 0 00 C G 0 00 C H 2900 3 5 8 3 1 300ml 300ml 500ml 0 3 psi P 0 3 psi 1 1 PT1 4 0 3 C 10 C 0 3 C 0 3 C 25 C 0 3 C 0 1 C 35 C 0 1 C 0 3 C 40 C 0 3 C 2 2 PT2 4 0 3 C 10 C 0 3 C 0 3 C 25 C 0 3 C 0 1...

Page 229: ...E S E T R A D I T I O N A L 4 4 1 CTU 4 2 CTU CTU 4 3 800 526 4455 5 5 8 30am 6 00pm 877 267 2314 24 866 840 9776 Arctic Sun 4 4 WEEE Bard A 115VAC 60Hz 4 5A 230VAC 50 60Hz 2 25A 100VAC 50 60Hz 5 0A...

Page 230: ...21 2 10 C 22 2 25 C 23 2 35 C 24 2 40 C 25 2 26 2 30 2 31 32 40 1 6 C 41 1 28 C 42 1 40 5 C 43 1 1 C 44 1 47 C 50 2 6 C 51 2 28 C 52 2 40 5 C 53 2 1 C 54 2 47 C 60 3 6 C 61 3 28 C 62 3 40 5 C 63 3 1...

Page 231: ...231 C H I N E S E T R A D I T I O N A L...

Page 232: ...232 TEMPERATURE MANAGEMENT SYSTEM C T U...

Page 233: ...2 234 1 3 234 1 4 235 1 5 235 2 235 2 1 235 2 2 235 3 236 3 1 236 3 2 CTU 237 3 3 237 3 4 237 3 5 237 3 5 1 237 3 5 2 237 3 5 3 237 3 5 4 238 3 5 5 238 3 5 6 238 3 5 7 238 3 5 8 238 4 239 4 1 239 4 2...

Page 234: ...T SYSTEM C T U 1 1 1 CTU Arctic Sun 5000 CTU CTU CTU CTU CTU Arctic Sun 5000 Arctic Sun CTU 1 2 CTU Arctic Sun CTU Medivance 1 3 CTU 1 1 CTU 3072519 ETL Arctic Sun CTU C22 2 61010 1 UL 61010 1 ETL Int...

Page 235: ...o 5 95 o 375mmHg 795mmHg o 375mmHg 795mmHg CTU IPX0 1 5 CTU CTU CTU 2 2 1 90 1 12 2 5 3 18 4 33 5 25 2 2 A CTU B PT1 5000 1 C PT2 5000 2 D TO TEMP OUT 5000 E Arctic Sun 5000 F G H Arctic Sun 5000 I CT...

Page 236: ...236 TEMPERATURE MANAGEMENT SYSTEM C T U 2 4 Arctic Sun 5000 2 5 CTU Arctic Sun 5000 2 6 CTU 1 2 7 CTU 2 3 2 8 CTU 3 3 1 CTU Arctic Sun 5000 5000 CTU...

Page 237: ...S1 A B 10 8 C C 25 5 C D 34 5 C E 40 0 C F CTU S2 A 1 2 B 1 CTU 2 C 1 S3 A B S4 250W AC AC CTU 3 4 PT1 PT2 TO o PT1 o PT2 o TO TO PT1 PT2 PT1 PT2 PT1 PT2 TO 3 5 CTU Arctic Sun 5000 3 5 1 3 5 2 1 10 C...

Page 238: ...CTU A H 0 CTU CTU CTU CTU CTU A 10 87 C B 25 54 C C 34 52 C D 39 98 C E 0 00 C F 0 00 C G 0 00 C H 2900 3 5 8 3 1 300ml min 300ml min 500ml min 0 3psi P 0 3psi 1 1 PT1 4 0 3 C 10 C 0 3 C 0 3 C 25 C 0...

Page 239: ...O R E A N 4 4 1 CTU 4 2 2 CTU CTU 4 3 800 526 4455 5 5 8 30 6 00 EST 877 267 2314 24 866 840 9776 Arctic Sun 4 4 WEEE Bard A 115VAC 60Hz 4 5A 230VAC 50 60Hz 2 25A 100VAC 50 60Hz 5 0A 6 3A Slow Blow 2...

Page 240: ...21 2 10 C 22 2 25 C 23 2 35 C 24 2 40 C 25 2 26 2 30 2 31 32 40 1 6 C 41 1 28 C 42 1 40 5 C 43 1 1 C 44 1 47 C 50 2 6 C 51 2 28 C 52 2 40 5 C 53 2 1 C 54 2 47 C 60 3 6 C 61 3 28 C 62 3 40 5 C 63 3 1...

Page 241: ...241 K O R E A N...

Page 242: ...A restricts this device to sale by or on the order of a physician Medivance Inc 321 South Taylor Avenue Suite 200 Louisville Colorado 80027 USA Phone 303 926 1917 Toll Free 800 526 4455 option 5 optio...

Reviews: