![AGFA Drystar 5302 Technical Documentation Manual Download Page 394](http://html.mh-extra.com/html/agfa/drystar-5302/drystar-5302_technical-documentation-manual_2866645394.webp)
Maintenance
DD+DIS149.05E
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 10
DRYSTAR 5302
Edition 1, Revision 0
Agfa Company Confidential
(Type
5366/100)
2005-09-09
4
Maintenance points - Power off
The maintenance with powered-off machine comprises following tasks:
Task
Details: see...
Check of condition of belts and gears
Cleaning of Power Supply Air Inlet
Cleaning of Thermal print head
4.1
Check of general condition
(1) Perform a visual check of
•
Power- and network cables.
•
Covers
•
The Output tray
•
The upper and lower Input tray.
(2) Disconnect the power and network cable
(3) Remove the left and right cover to prepare the printer for the next steps
(4) Remove the films in the film trays including white plastic sheet
4.2
Check of condition of belts and gears
(1) Check all belts and gears for visible wear
4.3
Visual check of the cables
(1) Check cables inside the printer: The insulation may not show any sign of
wear or damage.