![AGFA Drystar 5302 Technical Documentation Manual Download Page 395](http://html.mh-extra.com/html/agfa/drystar-5302/drystar-5302_technical-documentation-manual_2866645395.webp)
DD+DIS149.05E
Maintenance
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
DRYSTAR 5302
Chapter 9 / 11
4.4
Cleaning of the suction cups
(1) Pull out the upper pickup unit
2005-09-09 (Type
5366/100)
Agfa Company Confidential
536609da.cdr
1
(2) Clean the suction cups with a lint
free cloth lightly moistened with
Isopropyl alcohol.
(3) Insert the upper pickup unit
(4) Repeat this step for the lower
pickup unit
Figure 7
536609cc.cdr
2
536609db.cdr
3
Figure 8
Figure 9
4.5
Cleaning of Electronics Compartment
(1) Remove the EMI shielding
536609ac.cdr
(2) Clean the electronics
compartment with a vacuum
cleaner.
Figure 10