DD+DIS150.05E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR 5302
Chapter 3.6 / 6
05-2009 Type
5366/100
Agfa Company Confidential
Verification
(1) On the local keypad select key-operator menu '3 Print Image - 1 Testimage –
upper or lower tray - SMPTE_xxx.tif': An SMPTE test image will be printed.
(2) Measure the left and right blank frame in print direction: It must be
≥
1,5 mm.
Result
The horizontal film alignment procedure is performed.
2
Calibrations
Following calibrations have to be performed at the printer:
Adjustment
To be done:
Details see:
Film calibration
•
During
maintenance.
•
After replacement of the Thermal
print head (TPH).
•
During the troubleshooting
procedure.
Internal densitometer
(CDM) calibration
•
When the density reproduction
check indicates to calibrate the
CDM.
•
During the troubleshooting
procedure.
Print Head (TH)
calibration
•
During the troubleshooting
procedure.
•
During the TH cleaning procedure.
Refer to section 3.2 in this chapter.
Smart Profile
Calibration
•
During the troubleshooting
procedure.
•
In certain cases during
maintenance. Refer to section 9,
Maintenance.
Compliance Film
Calibration
•
May be required as last step of an
Afssaps / EUREF calibration
procedure.