![AGFA Drystar 5302 Technical Documentation Manual Download Page 170](http://html.mh-extra.com/html/agfa/drystar-5302/drystar-5302_technical-documentation-manual_2866645170.webp)
DD+DIS150.05E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.3 / 14 of 72
03-2011 Type
5366/100
Agfa Company Confidential
HC303004
: upperpickupmodule.filmpickup
Type
Accidental
Context
Action
None
Reason
Upper input jam
Cure
Remove upper input jam
HC303005
: upperpickupmodule.s18.low
Type
Accidental
Context
Action
None
Reason
Upper input jam
Cure
Remove upper input jam
HC303006
: upperpickupmodule.s18.high
Type
Accidental
Context
Action
None
Reason
Internal
jam
Cure
Remove internal jam
HC303007
: upperpickupmodule.pickup.skew
Type
Accidental
Context
Action
None
Reason
Upper input jam
Cure
Remove upper input jam