DD+DIS150.05E
Repair and Service
Software Releases and Patches
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Symptom 2
Electromagnetic radiation of the Themis CPU above specifications.
This symptom has no impact for the printer operation, violates however the specified
electromagnetic radiation values.
Cause
A certain component on the Themis CPU named "Spread Spectrum Clock (SSC)" is
switched off during operation.
4
Software Release 1.9.0_C2
Type of new Software*
Software Upgrade
Release Date
06/2007
Printer Serial Numbers
≥
SN 3930, plus 3642, 3645, 3711, 3714, 3798,
3803, 3815, 3856, 3862, 3864, 3865, 3867 to
3880, 3883, 3887, 3895, 3903 to 3909, 3911 to
3913, 3915 to 3924, 3926 to 3928
Availability
MedNet, GSO Library
Prerequisites
None
*NOTE:
A software upgrade is a functional extension. An update (software patch) is an
internal change designed to correct bugs, fix defects, or otherwise minor
improvements to performance without changing the software's basic design,
structure or functionality.
On MedNet, GSO Library always the latest available software release can be
downloaded.
4.1
New Features
Software 1.9.0_C2 contains following new features:
Support of RoHS conform printed circuit boards
Improvement of the Thermal Model, what improves constancy if several films are
printed short after each other
NOTE:
As of software 1.9.0_C2 the key-operator screens are available in English language
only.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.8 / 9
03-2011 Type
5366/100
Agfa Company Confidential