DD+DIS150.05E
Repair and Service
Electrical and mechanical codings, fuse tables and LED’s
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
Drystar 5302
Chapter 3.4 / 3
2005-09-09 (Type
5366/100)
Agfa Company Confidential
2 Fuses
The printer has no fuses that can be replaced.
If one of the fuses in the power supply is defective, the whole Power supply
unit must be replaced - refer to section 3.5.
3 LEDs
The following tables show the status of the LED indications and the
corresponding function:
LED's on...
Refer to
DEVIL Board
THEMIS Board
Power supply
Front Panel
Rear Panel
3.1 DEVIL
Board
Name Led
Color
Function
(**) NTRAFFIC
D1
yellow
Network
activity
(**) NGOOD_LINK
D2
yellow
Network
connected
(**) 100MBit_speed
D3
yellow
High
speed
network
IMGCTRL_0 D4
yellow
Software
controllable
D5
green
+3.3V
present
NOT present
D6
NOT present
D7
NOT present
D8
D9
green
+24V
present
D10
green
+12V
present
NOT present
D11
IMGCTRL_CONFDONE
D12 yellow
Off: FPGA config done.
ON: FPGA not
configured
IMGCTRL_1 D13
yellow
Software
controllable
UP_LED0 D14
yellow
Software
controllable
UP_LED1 D15
yellow
Software
controllable
(**) IDE_DASP
D16
yellow
IDE
activity
D17
green
-12V
present
D18
green
+5V
present
CP_ACTLED D19
yellow
Disc activity controlled
by PORTEX
Legend : (**) are LEDs situated at the rear panel (refer to 3.5).