DD+DIS150.05E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.3 / 62 of 72
03-2011 Type
5366/100
Agfa Company Confidential
HC311020
: inputsensormodule.working
Type
Defect
Context
11
Action
Replace
Reason
S4 bracket module is defective
Cure
Power off printer and replace
Part
MOD10
MODULE : camdrivemodule
Elabel : No
HC313000
: camdrivemodule.home
Type
Accidental
Context
Action
None
Reason
Internal
jam
Cure
Remove internal jam
HC313001
: camdrivemodule.m2.connection
Type
Defect
Context
E313001
Action
SeeManual
Reason1
Motor M2 not connected
Cure1
Connect motor M2
Reason2
Cam shaft module defect
Cure2
Replace
Part
MOD13