DD+DIS150.05E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.3 / 37 of 72
03-2011 Type
5366/100
Agfa Company Confidential
HC306020
: outputmodule.working
Type
Defect
Context
06
Action
Replace
Reason
Output module defect
Cure
Replace
Part
MOD06
HC306021
: outputmodule.m4.connection
Type
Defect
Context
E306021
Action
SeeManual
Reason1
Output motor not connected
Cure1
connect output motor
Reason2
Output module defect
Cure2
Replace
Part
MOD06
HC306022
: outputmodule.m4.shortcircuit
Type
Defect
Context
E306022
Action
SeeManual
Reason
Output module defect
Cure
Replace
Part
MOD06