![AGFA Drystar 5302 Technical Documentation Manual Download Page 88](http://html.mh-extra.com/html/agfa/drystar-5302/drystar-5302_technical-documentation-manual_2866645088.webp)
DD+DIS150.05E
Functional Description
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Errors are internally divided into:
•
Events
•
Warnings
•
Defects
In general events are user clearable, warnings are "key-operator"
(= experienced user or hospital technician) clearable, whereas defects have to be
repaired by service personnel.
Warnings and error messages are written into the infocounter.
Filmrun
Step 1
user
clearable
Errors: Classification and treatment
536602af.cdr
key-operator
clearable
service
clearable
Infocounter
Online
Diagnosis
event
Message in plain text.
E.g. 'close upper cover'
warning
Message in plain text:
E.g. CDM not calibrated
defect
Error message:
HC10907:
Suspension motor M3 not
connected or malfunctioning
everything
okay?
no
yes
Filmrun
Step n
B
Figure 22
Edition 1, Revision 1
DRYSTAR 5302
Chapter 2 / 24
06-2009 Type
5366/100
Agfa Company Confidentia