![AGFA Drystar 5302 Technical Documentation Manual Download Page 324](http://html.mh-extra.com/html/agfa/drystar-5302/drystar-5302_technical-documentation-manual_2866645324.webp)
DD+DIS150.05E
Repair and Service
Software Releases and Patches
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
12
Software Release 2.1.0_C3
Type of new Software*
Software Update
Release Date
02/2011
Printer Serial Numbers
SN
≥
13990 plus SN 13865
Availability
MedNet, GSO Library
Prerequisites
None
*NOTE:
A software upgrade is a functional extension. An update (software patch) is an
internal change designed to correct bugs, fix defects, or otherwise minor
improvements to performance without changing the software's basic design,
structure or functionality.
On MedNet, GSO Library always the latest available software release can be
downloaded.
12.1
New Feature of Software 2.1.0_C3
By installing software 2.1.0_C3, the Devius board is re-programmed to revision 9.
This is a solution for the warning message “RF Tag Problem” after 10 printed films of a
new film pack.
For details see DRYSTAR AXYS SB 15 / DRYSTAR 5302 SB 20, Software 2.1.0_C3
available to solve Warning Message “RF Tag Problem“ after 10 printed Films of a new
Film Pack, DD+DIS037.11E (
/
).
NOTE:
Software 2.1.0_C3 supports DRYSTAR 5302 and DRYSTAR AXYS.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.8 / 22
03-2011 Type
5366/100
Agfa Company Confidential