![AGFA Drystar 5302 Technical Documentation Manual Download Page 189](http://html.mh-extra.com/html/agfa/drystar-5302/drystar-5302_technical-documentation-manual_2866645189.webp)
DD+DIS150.05E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.3 / 33 of 72
03-2011 Type
5366/100
Agfa Company Confidential
MODULE : outputmodule
Elabel : Yes
HC306000
: outputmodule.connected
Type
Defect
Context
E306000
Action
SeeManual
Reason1
XS2 not connected
Cure1
Connect
XS2
Reason2
XS3 not connected
Cure2
Connect
XS3
HC306001
: outputmodule.cdm.init
Type
Failure
Context
W306001
Action
None
Reason
Output module defect
Cure
Replace
Part
MOD06
HC306002
: outputmodule.intensity.scan
Type
Accidental
Context
Action
None
Reason
Internal film jam
Cure
Remove film jam
HC306003
: outputmodule.outputsensor.hi
Type
Accidental
Context
Action
None
Reason
Output film jam
Cure
Remove film jam