DD+DIS150.05E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.3 / 7 of 72
03-2011 Type
5366/100
Agfa Company Confidential
HC301004
: uppertraymodule.s11
Type
Failure
Context
W301004
Action
None
Reason1
S11 not connected
Cure1
Connect
S11
Reason2
Upper tray module defect
Cure2
Replace
Part
MOD01
HC301005
: uppertraymodule.s24
Type
Failure
Context
W301005
Action
None
Reason1
S24 not connected
Cure1
Connect
S24
Reason2
Upper tray module defect
Cure2
Replace
Part
MOD01
HC301006
: uppertraymodule.takeoversensors.working
Type
Defect
Context
E301006
Action
SeeManual
Reason1
S11 and S24 not connected
Cure1
Connect S11 and S24
Reason2
Upper tray module defect
Cure2
Replace
Part
MOD01